Philips V60 Ventilators Recalled for Power Management Board Failure
Respironics is recalling 23,581 Philips V60 Ventilators because a solder connection failure on the power board may cause the blower to lose power or shut down unexpectedly, potentially leading to hypercarbia or hypoxemia.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II and involves a medical device defect that may lead to hypercarbia or hypoxemia, which are significant physiological injuries, meeting the criteria for a Severe rating.
Plain-English summary
Respironics California, LLC is recalling 23,581 units of Philips V60 Ventilators equipped with a Power Management PCBA part number 1055906. The recall is due to a solder connection failure on the Power Management printed circuit board assembly. This defect may cause the blower to lose power, spool down, and trigger a High Priority Check Vent alarm, or it may cause an intermittent connection that disrupts operation and triggers an unexpected shutdown without an alarm.
The affected ventilators were distributed across all 50 U.S. states, the District of Columbia, Puerto Rico, and the U.S. Virgin Islands, as well as numerous countries outside the United States. The specific serial numbers affected are 100002908 to 100027093, 100027096 to 100085389, 100085399 to 100109746, 100109868 to 201003390, and 201003392 to 201010952.
Failures of the power management board may lead to hypercarbia or hypoxemia. Users should check their device serial numbers against the affected ranges and contact Respironics for further instructions on corrective actions.
The recalled product
- Product
- Philips V60 Ventilators with Power Management PCBA part number 1055906
- Manufacturer
- Respironics California, LLC
- Category
- Medical Device — Ventilator
- Affected units
- 23,581
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Serial Numbers: 100002908 to 100027093
- 100027096 to 100085389
- 100085399 to 100109746
- 100109868 to 201003390
- 201003392 to 201010952
Distribution
Related recalls
Same manufacturer · Respironics California, LLC
See all →- SeverePhilips Respironics V680 Ventilator May Stop Operating Due to Electrical Fault
FDA (Devices) · 2022-04-27
- SeverePhilips Respironics V60 Plus Ventilator Recalled for Electrical Circuit Failure Risk
FDA (Devices) · 2022-04-27
- SeverePhilips Respironics V60 Ventilator Recalled Due to Electrical Circuit Defect
FDA (Devices) · 2022-04-27
- SeverePhilips Respironics V60 Plus Ventilators with Defective Adhesive Recalled
FDA (Devices) · 2022-03-09
- SeverePhilips V60 Ventilator Recall: Expired Adhesive May Cause Device Failure
FDA (Devices) · 2022-03-09
Same hazard · power failure
See all →- CriticalCarestation 750 anesthesia systems recalled over potential unexpected shutdown
FDA (Devices) · 2025-12-24
- CriticalAnesthesia systems recalled over potential unexpected shutdown on mains power loss
FDA (Devices) · 2025-12-24
- CriticalAnesthesia power management boards recalled over potential unexpected system shutdown
FDA (Devices) · 2025-12-24
- HighBD FACSLyric Flow Cytometer Power Supply Module Failure Risk
FDA (Devices) · 2025-01-01
- HighBD FACSLyric 2L4C Flow Cytometer Power Supply Failure Recall
FDA (Devices) · 2025-01-01
Similar recalls
- SeverePhilips Respironics V60 and V60 Plus Ventilators Power Management Failure Recall
FDA (Devices) · 2023-10-04
- SevereRespironics V60 Ventilator Recalled for Power Management Board Software Issue
FDA (Devices) · 2013-06-26
- HighPhilips Respironics V60 Ventilator Recalled for Backup Battery Failure Risk
FDA (Devices) · 2021-02-03
- HighPhilips Respironics V60 Ventilator Recall for Backup Battery Failure Risk
FDA (Devices) · 2021-02-03
- HighPhilips Respironics V60 Ventilator Recalled for Potential Backup Battery Failures
FDA (Devices) · 2021-02-03
- SeverePhilips Respironics V60 Ventilator Recalled Due to Electrical Circuit Defect
FDA (Devices) · 2022-04-27