Haemonetics Recalls TEG Manager Software Due to Alert Display Defect
Haemonetics is recalling TEG Manager software versions 1.1.0 through 4.1.1 because a defect causes incorrect out-of-range alerts for coagulation test results.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the defect involves a software display error that could lead to misinterpretation of test results, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Haemonetics Corporation is recalling TEG Manager software versions 1.1.0, 3.0.0, 4.0.0, 4.1.0, and 4.1.1, which are used in conjunction with the TEG 5000 device for in vitro coagulation studies. The recall involves 88 units distributed nationwide in the United States.
The recall is due to a software defect that impacts the displayed alerts for out-of-range test results. Specifically, reference range values received from the TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number. This rounding error can lead the software to display an out-of-range alert when the test result is actually in range, or fail to display an alert when the result is out of range.
Healthcare facilities using these specific software versions should contact Haemonetics Corporation for instructions on how to resolve the defect.
The recalled product
- Product
- TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies
- Manufacturer
- Haemonetics Corporation
- Category
- Medical Device — Software
- Affected units
- 88
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0
- 3.0.0
- 4.0.0
- 4.1.0
- and 4.1.1.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Haemonetics Corporation
See all →- HighTEG5000 Analyzer displays incorrect PlateletMapping assay results in TEG Manager
FDA (Devices) · 2022-07-06
- HighFDA Recalls TEG 5000 Fibrinogen Reagent Kits Due to Outdated Reference Ranges
FDA (Devices) · 2022-07-06
- HighHaemonetics Recalls Cell Saver Bowl Sets Due to Leak Risk
FDA (Devices) · 2020-02-05
- HighHaemonetics Recalls Cell Saver 5/5+ Bowl Kits Due to Leak Risk
FDA (Devices) · 2020-02-05
- HighHaemonetics Recalls CS5/5+ Bowl Kits Due to Potential Leaks
FDA (Devices) · 2020-02-05
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25