ModerateFDA (Devices)·Z-1598-2013·Announced 2013-07-17

Toshiba Kalare Diagnostic X-Ray System Recalled for Image Quality Defect

Toshiba American Medical Systems is recalling 153 Kalare Diagnostic X-Ray Systems because the large focus heats up unexpectedly, causing brighter images than expected.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is a voluntary precautionary action by the firm for a functional defect that affects image quality (brightness) but does not involve reported injuries, fatalities, or high-risk pathogens.

Plain-English summary

Toshiba American Medical Systems Inc is recalling 153 units of the TOSHIBA Kalare Diagnostic X-Ray System (K110785) distributed nationwide in the United States. The recall was initiated by the firm due to a defect in the equipment's operation.

The issue occurs when radiographic conditions with the small focus are set and the radiography ready switch is held down while using the Automatic Exposure Function (AEC). In these instances, the large focus is also heated after a few seconds or more. This causes a radiographic tube current exceeding the set value to flow, resulting in a brighter image than expected being acquired.

The source text does not specify instructions for consumers or healthcare providers on how to resolve the issue, nor does it report any injuries or illnesses associated with the defect.

The recalled product

Product
TOSHIBA Kalare Diagnostic X-Ray System K110785
Affected units
153

Distribution

Distributed nationwide across the United States.