The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-1552-2020·Announced 2020-04-01

Quidel Triage TOX Drug Screen Control 1 Kit Recall for Chip Error

Quidel is recalling Triage TOX Drug Screen Control 1 kits due to a programming error in the control chip that causes false failing results for AMP and mAMP tests.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a software/programming error causing false failing results, which is a moderate risk of diagnostic error without reported injuries or deaths.

Plain-English summary

Quidel Cardiovascular Inc. is recalling 123 kits of the Triage TOX Drug Screen Control 1. The affected product consists of boxes containing five Control 1 vials. The recall is limited to Lot Number C3470AN and Part Number 94413.

The recall was initiated due to a programming error in the Control CODE CHIP module for this specific lot. This error causes the Triage Meter to display a failing control result for the AMP and mAMP tests, even when the control did not actually fail. This issue results in inappropriate failing results being displayed for these specific tests.

The affected kits were distributed in the United States to the states of Idaho, Florida, Georgia, Illinois, Texas, Kentucky, Missouri, Alabama, Utah, Iowa, Washington, Ohio, Arizona, New York, Oklahoma, Oregon, Indiana, and Kansas. They were also distributed to California and the Netherlands. Users should stop using the affected kits and contact Quidel for further instructions.

The recalled product

Product
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
Affected units
123

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot Number: C3470AN Part Number 94413

Distribution

Same manufacturer · QUIDEL CARDIOVASCULAR INC