Beckman Coulter Recalls MicroScan Dried Gram Negative MIC Panels
Beckman Coulter is recalling MicroScan Dried Gram Negative MIC/Combo Panels because a technical bulletin failed to mention performance limitations for Proteus mirabilis and Imipenem.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a temporary or medically reversible adverse health consequence. The hazard is a performance limitation in a diagnostic tool rather than a direct physical injury or contamination, fitting the Moderate severity criteria for labeling or performance errors without reported harm.
Plain-English summary
Beckman Coulter, Inc. is recalling MicroScan Microbiology Systems Dried Gram Negative MIC/Combo Panels, specifically the Neg/Urine Combo 61 panel (Catalog B1017-414). These panels are used in laboratories to determine antimicrobial agent susceptibility and identify aerobic and facultatively anaerobic gram-negative bacilli.
The recall was initiated because a Technical Support Bulletin issued in February 2013 did not mention a performance limitation regarding Proteus mirabilis and Imipenem when using the CLSI M100-S20 revised interpretive criteria. This omission could affect the accuracy of susceptibility reporting for these specific conditions.
Approximately 216,240 boxes were distributed in the United States and 84,249 boxes were distributed outside the U.S. to various countries including Canada, Germany, and China. Healthcare facilities and laboratories that use these panels should contact Beckman Coulter for further instructions on handling the affected product.
The recalled product
- Product
- MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 61, Catalog B1017-414 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and…
- Manufacturer
- Beckman Coulter, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Catalog number: B1017-414
- Siemens Material Number (SMN): 10483097.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 7 recalled by Beckman Coulter, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 7 products in this recall →Related recalls
Same manufacturer · Beckman Coulter, Inc.
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- High
Same hazard · performance limitation
See all →- ModerateBeckman Coulter MicroScan Panels Recalled for Missing Proteus Performance Warning
FDA (Devices) · 2015-05-06
- ModerateBeckman Coulter Recalls MicroScan Dried Gram Negative MIC Panels
FDA (Devices) · 2015-05-06
- ModerateBeckman Coulter Recalls MicroScan Microbiology Panels Due to Performance Limitation
FDA (Devices) · 2015-05-06
Similar recalls
- ModerateBeckman Coulter Recalls MicroScan Neg Breakpoint Combo Panel Type 34
FDA (Devices) · 2015-05-06
- HighBeckman Coulter MicroScan gram-negative panels may contain contaminated wells
FDA (Devices) · 2026-04-01
- HighBeckman Coulter MicroScan Pos Combo Panels Recalled for False Negative Results
FDA (Devices) · 2016-01-06
- ModerateBeckman Coulter Recalls MicroScan Pos Breakpoint Combo Panels Due to False Negatives
FDA (Devices) · 2016-01-06
- HighBeckman Coulter MicroScan Pos Combo Panel Recall for False Negatives
FDA (Devices) · 2016-01-06
- SevereBeckman Coulter MicroScan Pos Breakpoint Combo Panel Recall
FDA (Devices) · 2016-01-06