The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-1531-2013·Announced 2013-06-19

International Biomedical Recalls VIA Medical Collection Bags Due to Connector Issues

International Biomedical is recalling VIA Medical Collection Bags because the connector can become occluded if the luer lock is over-tightened.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a situation where use of the product may cause temporary or medically reversible health consequences. The hazard involves potential occlusion of the connector, which is a functional defect rather than a direct injury or contamination risk, fitting the Moderate severity criteria for low-risk mechanical issues.

Plain-English summary

International Biomedical, Ltd. is recalling approximately 7,000 VIA Medical Collection Bags. These bags are accessory waste bags used with the sterile LVM blood gas sensor and LVM Monitor. The affected products are packaged in tyvek pouches and are included in replacement kit 777-3103 or LVM sensor kit 777-3102. The collection bags have model/part number 209-2113 and specific lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, and 4489.

The recall was initiated after the company received reports of interference between the collection bag and the extension set connectors. Evaluation of the components indicated that the collection bag connector can become occluded due to the over-tightening of the luer lock by the end user. To address this issue, the connector design on the collection bag has been improved.

The affected products were distributed nationwide. Users should check their inventory for the specific model and lot numbers listed above. If the product is found, it should be removed from use and returned to the manufacturer or a designated distributor for replacement or refund, in accordance with the recall instructions provided by International Biomedical.

The recalled product

Product
VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit…
Affected units
7,000

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (8)

  • Model/Part Number: 209-2113
  • or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389
  • 4465
  • 4478
  • 4516
  • 4530
  • 4430
  • 4489

Distribution

Distributed nationwide across the United States.

Similar recalls