The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-1457-2015·Announced 2015-04-29

Siemens ADVIA Chemistry Lipase Reagent Recalled for Carryover Contamination

Siemens Healthcare Diagnostics is recalling ADVIA Chemistry Lipase Reagent because carryover from other reagents can cause falsely elevated test results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II. Per the rubric, FDA Class II recalls are assigned a score of 4 (Severe), particularly when they involve significant diagnostic errors that could impact patient care.

Plain-English summary

Siemens Healthcare Diagnostics, Inc. is recalling ADVIA Chemistry Lipase Reagent, Catalog Number B01-4840-01. The recall affects 7,814 units with Lot Numbers 318706 and 332463, which have expiration dates of June 30, 2015, and August 31, 2015. The product was distributed nationwide in the United States and to numerous countries worldwide.

A Siemens internal investigation confirmed that the reagent's current contamination avoidance settings do not prevent carryover. This carryover can result in falsely elevated ADVIA Chemistry Lipase outlier results when the reagent is run alongside Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.

This recall is classified as Class II by the FDA. Users should stop using the affected reagent and contact Siemens Healthcare Diagnostics for instructions on return or replacement.

The recalled product

Product
ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.
Affected units
7,814

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (6)

  • Catalog Number: B01-4840-01
  • Material Number (SMN)10311896
  • Lot Numbers: 318706
  • 332463
  • expiration dates 2015-06-30
  • 2015-08-31

Distribution

Distributed nationwide across the United States.