Siemens COHERENCE Oncologist Software Recall for Incorrect Dose Positioning
Siemens is recalling COHERENCE Oncologist software due to a calculation error that could lead to incorrect patient repositioning and radiation dose delivery to the wrong location.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as Severe because it involves a medical device defect that could result in significant injury through incorrect radiation dose delivery to the wrong location, consistent with the rubric for significant injury reports or high-risk device defects.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling the COHERENCE Oncologist software, Model Nos. 07333680 & 07351898. The recall involves software versions 2.0.49, 2.0.50, and 2.0.51. A total of 64 units were distributed worldwide, with 26 units distributed domestically in the United States.
The recall is issued because a potential safety issue arises when an offset calculation is followed by a filter operation. Correcting the image alignment after this specific sequence of operations results in incorrect offset values. This error could lead to the incorrect repositioning of the patient and the delivery of radiation dose to the wrong location.
The affected software was distributed domestically to California, Florida, Georgia, Iowa, Massachusetts, Michigan, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Dakota, Tennessee, and West Virginia. It was also distributed to foreign locations worldwide. Users of this software should contact Siemens Medical Solutions USA, Inc. for further instructions on correcting the issue.
The recalled product
- Product
- COHERENCE Oncologist, Model Nos. 07333680 & 07351898
- Manufacturer
- Siemens Medical Solutions USA, Inc
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Software versions 2.0.49
- 2.0.50
- 2.0.51
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · incorrect dose delivery
See all →- SeverePTW Freiburg Recalls VERIQA RT MonteCarlo 3D Software Due to Bolus Display Error
FDA (Devices) · 2026-09-23
- HighAccuray TomoTherapy System Recalled for MLC Tickling Error
FDA (Devices) · 2021-01-13
- HighMevion S250 Proton Therapy System Recalled for Software Defect
FDA (Devices) · 2014-12-17
- HighElekta Monaco System Recalled for Incorrect Radiation Treatment Plan Delivery
FDA (Devices) · 2014-09-10
Similar recalls
- ModerateSiemens Medical Solutions Recalls COHERENCE RT Therapist Software for Safety Issues
FDA (Devices) · 2013-04-10
- ModerateSiemens COHERENCE RT Therapist Software Recalled for Performance Defects
FDA (Devices) · 2013-04-10
- ModerateSiemens COHERENCE RT Therapist Software Recall for Performance Defects
FDA (Devices) · 2013-04-10
- ModerateSiemens COHERENCE RT Therapist Software Recall for Safety Defects
FDA (Devices) · 2013-04-10
- SevereSiemens Syngo.via RT Image Suite Software Recall for Calibration Offset Error
FDA (Devices) · 2020-09-30
- ModerateSiemens Medical Solutions Recalls COHERENCE RT Therapist Software for Performance Defects
FDA (Devices) · 2013-04-10