Roche Recalls Elecsys BRAHMS PCT Immunoassay Due to Decreased Sample Recovery
Roche Diagnostics is recalling 47 units of the Elecsys BRAHMS PCT Immunoassay because it shows decreased recovery of patient samples on the cobas e 801 module compared to other analyzers.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a situation where use of the product may cause temporary or medically reversible health consequences. The source text describes a performance defect (decreased recovery) without reporting any specific injuries, illnesses, or fatalities, aligning with a Moderate severity rating for a diagnostic device.
Plain-English summary
Roche Diagnostics Corporation is recalling 47 units of the Elecsys BRAHMS PCT Immunoassay. This product is used for the in vitro quantitative determination of procalcitonin (PCT) in human serum and plasma. The recall is limited to Lot number 204084.
The recall was initiated because the product exhibits a decreased recovery of patient samples when used on the cobas e 801 module. This performance issue was observed in comparison to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.
The affected units were distributed in the United States to the states of California, Florida, and Indiana. Healthcare facilities and laboratories that have received this specific lot number should stop using the product and contact Roche Diagnostics for further instructions.
The recalled product
- Product
- Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).
- Manufacturer
- Roche Diagnostics Corporation
- Hazard
- Affected units
- 47
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Elecsys BRAHMS PCT: 07301715190 Lot number 204084
Distribution
Related recalls
Same manufacturer · Roche Diagnostics Corporation
See all →- ModerateRoche Tina-quant lgA Gen.2 Assay Recalled Due to Reagent Contamination
FDA (Devices) · 2019-07-24
- LowRoche cobas p 701 Post-Analytical Units Recalled for Loose Flap
FDA (Devices) · 2019-07-10
- LowRoche cobas p 501 Post-Analytical Units Recalled for Loose Flap
FDA (Devices) · 2019-07-10
- HighRoche Cobas infinity Software Recall for Missing Quality Control Alarms
FDA (Devices) · 2019-05-29
- SevereRoche Recalls cobas e 801 Immunoassay Analyzers Due to Signal Drops
FDA (Devices) · 2019-05-08
Same hazard · inaccurate results
See all →- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
- Highepoc BGEM BUN Test Card for epoc Blood Analysis System
FDA (Devices) · 2026-05-06
- HighAbbott i-STAT EG7+ Blood Gas Cartridges Reporting Inaccurate Results
FDA (Devices) · 2026-05-06