Stryker Irrigator Tips Recalled for Incorrect Sterilization Temperature Instructions
Stryker Endoscopy is recalling reusable irrigator tips because the instructions listed incorrect pre-vacuum steam sterilization temperatures.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the hazard is a minor labeling error regarding sterilization temperatures with no reported injuries or illnesses, fitting the criteria for a moderate severity score.
Plain-English summary
Stryker Endoscopy is recalling Stryker Irrigator Reusable Tips, specifically the 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. These suction irrigator tips are used to aid in cleaning and clearing the operative site during laparoscopic surgery. The recall covers 147,921 units of all sizes, distributed worldwide including the United States, Canada, Mexico, and various countries in Europe, Asia, and South America.
The recall was initiated because the Instructions for Use incorrectly listed the pre-vacuum steam sterilization (wrapped method) temperature as 121-123°C, instead of the correct 132-133°C. This labeling error could potentially affect the sterilization process if followed strictly.
Healthcare facilities and users of these devices should stop using the affected tips and contact Stryker Endoscopy for further instructions on how to proceed with the recall.
The recalled product
- Product
- Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Manufacturer
- Stryker Endoscopy
- Hazard
- Affected units
- 147,921
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Manufacturer Part Number: 250-070-472
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 20 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 20 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · mis labeling
See all →- SevereImperio Distribution Recalls Beef Jerky Produced Without Federal Inspection
USDA FSIS · 2026-10-24
- HighJ K Patisserie Chocolate Chip Cookies Recalled for Undeclared Soy
FDA (Food) · 2026-10-14
- HighFreshPoint Recalls Cranberry Chicken Salad Wedges for Undeclared Egg Allergen
FDA (Food) · 2026-10-14
- HighJ K Patisserie & Bakery LLC Recalls Croissants for Undeclared Egg
FDA (Food) · 2026-10-14
- LowMindray ePM Series Patient Monitor Manuals Recalled for Alarm Clarification
FDA (Devices) · 2026-10-14
Similar recalls
- SevereStryker Recalls Disposable StrykeFlow Tips Due to Sterility Concerns
FDA (Devices) · 2014-07-30
- ModerateStryker 3.5mm Trocar Recalled for Sterilization Instructions Error
FDA (Devices) · 2014-12-24
- ModerateStryker Laparoscopic Scissors Recalled for Sterilization Instructions Error
FDA (Devices) · 2014-12-24
- ModerateStryker Laparoscopic Reducers Recalled for Sterilization Parameter Errors
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors
FDA (Devices) · 2014-12-24
- ModerateStryker Laparoscopic Dissector Recall Due to Sterilization Documentation Error
FDA (Devices) · 2014-12-24