Medica Wash Kit Recalled for Inconsistent Lipase Test Results
Medica Corporation is recalling Wash Kit REF 10680 because an ineffective wash solution causes inconsistent lipase test results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the hazard involves diagnostic inaccuracy (inconsistent values) rather than direct physical injury or death, fitting the criteria for a moderate severity recall involving low-risk contamination or functional defects.
Plain-English summary
Medica Corporation has initiated a field recall of the Wash Kit (REF 10680), specifically Lots 16279, 17045, and 17215. The product is a wash solution used with the Medica EasyRA clinical chemistry analyzer. It is applied prior to a lipase (LIP) assay when a triglyceride (TRIG) assay is also part of the work list.
The recall was initiated due to reported inconsistent lipase values when analyzing patient samples. The firm's internal investigation identified the cause as an ineffective wash solution, which results in interference carryover from the triglyceride assay when both tests are run on the same work list.
The affected units were distributed worldwide and in the US. A total of 392 units are involved in this recall. Users of the Medica EasyRA analyzer should check their inventory for these specific lot numbers and stop using the affected wash kits.
The recalled product
- Product
- Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of…
- Manufacturer
- Medica Corporation
- Affected units
- 392
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Lots 16279
- 17045 and 17215
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Medica Corporation
See all →- HighCapillary tubes recalled over a positive bias in potassium results
FDA (Devices) · 2026-02-25
- ModerateMedica Wash1 Solution Recalled for Ineffective Probe Cleaning
FDA (Devices) · 2019-05-15
- HighMedica EasyCAL HbA1c Calibrator Recalled for Potential Result Deviations
FDA (Devices) · 2018-04-04
Same hazard · diagnostic inaccuracy
See all →- High
- HighAtellica CH Enzymatic Hemoglobin A1c assay well-to-well bias
FDA (Devices) · 2026-07-15
- HighTBS iNsight bone density software calculation variability on Hologic machines
FDA (Devices) · 2025-03-12
- HighBD PCR Cartridges for MAX System signal drift in select lots
FDA (Devices) · 2025-01-29
- HighHPV-16 Diagnostic Probe Recalled for Suboptimal DNA Hybridization
FDA (Devices) · 2024-12-04
Similar recalls
- ModerateMedica Wash1 Solution Recalled for Ineffective Probe Cleaning
FDA (Devices) · 2019-05-15
- ModerateAbbott Lipase Reagent Recalled for Incorrect Package Insert Parameters
FDA (Devices) · 2014-11-05
- ModerateSiemens ADVIA Chemistry Lipase Recalled for Calibration Failures
FDA (Devices) · 2020-03-04
- ModerateRandox Lipase Test System Recalled for Carryover Interference
FDA (Devices) · 2016-10-26
- ModerateRandox Lipase Assay Reagent Recalled for Imprecision in Quality Control and Patient Samples
FDA (Devices) · 2020-08-12
- ModerateRandox Lipase Assay Recalled for Potential Contamination on Rx Instruments
FDA (Devices) · 2018-05-16