SR Instruments FG3105 Patient Scales Recalled Due to Load Cell Failures
SR Instruments is recalling 806 FG3105 patient scales used with Joerns Healthcare lifters after two load cell failures caused patient falls.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a structural defect in a medical device that resulted in reported patient falls, meeting the criteria for 'significant injury/property-damage reports' and 'structural vehicle defect' analogs in the rubric.
Plain-English summary
SR Instruments Inc. is implementing a field correction for 806 FG3105 Patient Scales. These scales are designed for use with Joerns Healthcare Patient Lifters and are connected between the patient hoist mechanism and the patient sling apparatus. The affected units are identified by serial numbers 1005 through 1780, 1785 through 1809, 1866 through 1868, and 1870 through 1875.
The recall was initiated after the company received two incident reports regarding field failures of a tensile type load cell. These mechanical failures resulted in patient falls. The affected products were distributed nationwide in the United States, specifically within the state of Texas.
Consumers and healthcare facilities should stop using the affected scales immediately. The scale is not intended for patient transport. Users should contact SR Instruments Inc. for further instructions on the field correction.
The recalled product
- Product
- SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected betwe
- Manufacturer
- Sr Instruments Inc
- Affected units
- 806
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Serial numbers: 1005 thru 1780
- 1785 thru 1809
- 1866 thru 1868 and 1870 thru 1875.
Distribution
Related recalls
Same hazard · fall injury
See all →- HighBritax Recalls Infant Car Seats Due to Handle Detachment Risk
NHTSA · 2026-08-20
- HighQuantum Rehab Q6 Edge 2.0 Power Wheelchair Recall
FDA (Devices) · 2026-08-12
- HighPowered Wheelchair Controller Software May Cause Unexpected Brake Stop
FDA (Devices) · 2026-08-12
- HighPowered wheelchair controller recalled for unexpected braking hazard
FDA (Devices) · 2026-08-12
- SeverePowered Wheelchair Controller Software Recall by Pride Mobility
FDA (Devices) · 2026-08-12