The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-1099-2017·Announced 2017-02-01

Roche Recalls Accu-Chek Connect App Due to Insulin Dosing Errors

Roche Diabetes Care is recalling the Accu-Chek Connect app because software defects may cause incorrect insulin bolus recommendations, potentially leading to hypoglycemia.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II and the source text explicitly states that incorrect bolus advice could lead to serious health consequences such as hypoglycemia, meeting the criteria for a Severe rating.

Plain-English summary

Roche Diabetes Care, Inc. is recalling the Accu-Chek Connect Diabetes Management App for iOS and Android. The recall is due to software defects that may cause the app to disregard historical bolus data or fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement. These issues may result in incorrect bolus insulin recommendations being provided to users.

The affected app versions are distributed directly to end users via the Apple App Store and the Google Play App Store. The recall covers 27,243 units. The software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect.

Depending on the individual's metabolic situation, potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Users should not use the app to manually calculate a bolus. Consumers should stop using the affected app versions and follow instructions from the manufacturer or healthcare provider.

The recalled product

Product
Accu-Chek Connect Diabetes Management App
Affected units
27,243

Distribution

Distributed nationwide across the United States.

Similar recalls