Agency classClass IOur severity:SevereFDA (Devices)·Z-1060-2016·Announced 2016-03-16

Arrow UltraFlex IAB Recall for Potential Sheath Separation and Bleeding Risk

Arrow International is recalling UltraFlex IAB devices because the sheath body may separate from the hub, creating a potential for bleeding.

The agency’s classification

FDA (Devices): Class I

Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device."

Plain-English summary

Arrow International, Inc., Division of Teleflex Medical Inc., is recalling the UltraFlex IAB (7.5Fr 40cc, Product Code: IAB-06840-U). This medical device is utilized for intra aortic balloon counterpulsation therapy in the aorta, where balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.

The recall is being conducted because the sheath body may become separated from the sheath hub. If this separation occurs, there is a potential for bleeding from the device.

The recall affects 13,405 units in the US and 33,735 units outside the US, with worldwide distribution including the US (Nationwide) and numerous international countries. Specific batch codes are listed in the agency's structured data. Consumers and healthcare providers should stop using the affected devices and contact the manufacturer for further instructions.

The recalled product

Product
UltraFlex IAB: 7.5Fr 40cc; Product Code: IAB-06840-U The Arrow¿ IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases…

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 9 recalled by Arrow International, Inc., Division of Teleflex Medical Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 9 products in this recall →

Same manufacturer · Arrow International, Inc., Division of Teleflex Medical Inc.

See all →

Similar recalls