Allura Xper FD20 systems recalled over a missing cooling unit drip tray

Philips Medical Systems Nederland is recalling 79 units of the Allura Xper FD20, model 722012, because the drip tray beneath the cooling unit may not have been installed.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: Scored 2 because the record's stated consequence is a system shutdown, with no patient harm or reported injury.

Plain-English summary

The Philips Allura Xper FD20, model number 722012, is being recalled by Philips Medical Systems Nederland B.V.

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. The firm states this could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

The record lists 79 units under model number 722012, UDI-DI 00884838059054, and affected serial numbers listed in the enforcement record. Distribution was worldwide, including nationwide in the United States and to a wide range of other countries. The FDA has classified the action as Class II.

The stated consequence in the enforcement record is system shutdown, and no injuries are reported. Facilities should follow Philips' instructions.

The recalled product

Product
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
Manufacturer
Canon Medical System, USA, INC.
Hazard
  • missing-component
  • coolant-leak
  • short-circuit
  • system-shutdown

Distribution

Distributed nationwide across the United States.