Aomori Olympus Endoscopic Injector Recall Due to Package Seal Defects
Aomori Olympus is recalling specific endoscopic injectors because inspection found some sterile packages have inadequate seals, posing a risk of infection if non-sterile equipment is used.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (infection from non-sterile equipment) where no specific injuries or illnesses have been reported in the source text, aligning with the 'High' severity criteria for theoretical or unreported harm risks.
Plain-English summary
Aomori Olympus Co., Ltd. is recalling the INJECTOR FORCEMAX UP 2.8MMCH 25G 5MM (model no. NM-400L-0525). This medical device is used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
The recall was initiated after inspection confirmed that a small percentage of sterile packages have inadequate package seals, specifically partial heat seals. The use of non-sterile equipment presents a risk that patients may develop infections.
The recall affects 260,395 units distributed nationwide in the United States. Affected units are identified by the following lot codes: 01K, 04K, 05K, 07K, 09K, 98K, 99K, 9XK, 9YK, 9ZK, 05V, 06V, 07V, 08V, 09V, and 0XV. Healthcare providers and consumers should check their inventory for these specific lot codes and stop using the affected devices.
The recalled product
- Product
- INJECTOR FORCEMAX UP 2.8MMCH 25G 5MM, model no. NM-400L-0525 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
- Manufacturer
- Aomori Olympus Co., Ltd.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (16)
- 01K
- 04K
- 05K
- 07K
- 09K
- 98K
- 99K
- 9XK
- 9YK
- 9ZK
- 05V
- 06V
- 07V
- 08V
- 09V
- 0XV
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 111 recalled by Aomori Olympus Co., Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 111 products in this recall →Related recalls
Same manufacturer · Aomori Olympus Co., Ltd.
See all →- HighReusable Endoscopy Distal Attachments Recalled for Excessive DEHP Levels
FDA (Devices) · 2023-09-13
- HighOlympus Distal Attachment Recalled for Phthalate Exceeding Permissible Levels
FDA (Devices) · 2023-09-13
- HighAomori Olympus Recalls KD-612U Electrosurgical Snare Due to Seal Defects
FDA (Devices) · 2021-02-17
- HighAomori Olympus Recalls Aspiration Needles Due to Inadequate Package Seals
FDA (Devices) · 2021-02-17
- HighAomori Olympus Recalls 22G EZSHOT Aspiration Needles Due to Seal Defects
FDA (Devices) · 2021-02-17
Same hazard · infection
See all →- HighArrowhead DE LLC Recalls Arrow-LOK Implants Due to Sterility Assurance Concerns
FDA (Devices) · 2026-08-26
- SevereRegeneron Recalls Libtayo Injection Due to Sterility Assurance Concerns
FDA (Drugs) · 2026-08-26
- HighSurgipro Sutures Recalled for Potential Sterile Barrier Perforation
FDA (Devices) · 2026-08-19
- SevereSpectra Medical Devices Recalls Sodium Chloride Injection Ampules Due to Sterility Concerns
FDA (Devices) · 2026-08-19
- HighV-Loc 90 Absorbable Wound Closure Devices Recalled for Packaging Defects
FDA (Devices) · 2026-08-19