The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-1001-2015·Announced 2015-01-28

Genetic Testing Institute Recalls PF4 Concentrated Wash Reagent Lots

Genetic Testing Institute is recalling specific lots of PF4 Concentrated Wash (10x) because the reagent is slightly more dilute than intended, which may cause false positive test results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves potential false positive results due to reagent dilution, which is a quality issue without reported injuries or illnesses, fitting the Moderate severity criteria for low-risk contamination or labeling/quality errors.

Plain-English summary

Genetic Testing Institute, Inc. is recalling specific lots of PF4 Concentrated Wash (10x), Reference 403624. This reagent is packaged in ImmucorGamma LIFECODES PF4 IgG and PF4 Enhanced assay kits. The recall covers 6,206 kits distributed nationwide in the United States and internationally to Canada, Japan, Belgium, the European Union, Australia, Taiwan, Thailand, and Scotland.

The recall was initiated because the affected lots of PF4 Concentrated Wash are a slightly more dilute concentration than intended. This dilution may lead to slightly increased Optical Density (OD) values, causing samples with higher background reactivity to demonstrate false positive results (borderline reactivity) at or near the assay cutoff.

The affected reagent lot numbers are 3001620, 3001987, 3002083, and 3002153. Users of these assay kits should check their inventory for these specific lot numbers and contact the manufacturer for further instructions.

The recalled product

Product
PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1. ImmucorGamma LIFECODES PF4 IgG (catalog numbers HAT13G, HAT45G) and 2. ImmucorGamma LIFECODES PF4 Enhanced (catalog numbers X-HAT13, X-HAT45) The PF4 IgG assay is an Enzyme Linked…
Affected units
6,206

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (4)

  • Reagent Lot Numbers: 3001620
  • 3001987
  • 3002083
  • 3002153

Distribution

Distributed nationwide across the United States.

The notice also names 45 states specifically:

Same manufacturer · Genetic Testing Institute,inc

See all →

Similar recalls