Magstim Recalls Neurosign 100 Intraoperative Nerve Monitors for Voltage Configuration Error
The Magstim Company Limited is recalling seven Neurosign 100 Intraoperative Nerve Monitors because they were configured for 240V power instead of the required 120V for the US.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This is an FDA Class II recall involving a configuration error that poses a risk of electrical malfunction or damage, but the source text does not report any injuries, illnesses, or deaths.
Plain-English summary
The Magstim Company Limited is recalling seven units of the Neurosign 100 Intraoperative Nerve Monitor, Product Number 9883-01. These devices are indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve, during ENT surgical procedures.
The recall is due to a configuration error where the devices were dispatched with the onboard voltage selector set to 240V (European Standard) and equipped with 2xT315mAL fuses, rather than the required 120V setting and 2xT630mAL fuses for the United States and Canada. The affected units have serial numbers 2507, 2511, 2514, 2515, 2522, 2527, 2533, 2534, and 2537.
The affected devices were distributed to facilities in Minnesota, North Carolina, and New York. Users should stop using the affected devices and contact The Magstim Company Limited for instructions on how to proceed.
The recalled product
- Product
- Neurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.
- Manufacturer
- The Magstim Company Limited
- Hazard
- Affected units
- 7
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- Serial Numbers: 2507
- 2511
- 2514
- 2515
- 2522
- 2527
- 2533
- 2534
- 2537
Distribution
Related recalls
Same manufacturer · The Magstim Company Limited
See all →- HighMagstim Recalls Neurosign V4 Intraoperative Nerve Monitors Due to Cable Damage Risk
FDA (Devices) · 2020-09-02
- HighMagstim Recalls Rapid2 Stim Interconnecting Cables Due to Potential Failure
FDA (Devices) · 2020-05-20
- HighMagstim Horizon PSU Stim Interconnecting Cable Recalled for Potential Failure
FDA (Devices) · 2020-05-20
- HighMagstim HORIZON Ez Arm Recalled for Protruding Stainless-Steel Cable Strand
FDA (Devices) · 2019-02-20
- LowMagstim Rapid Therapy System Recalled for Missing User Manuals
FDA (Devices) · 2018-04-18
Same hazard · electrical malfunction
See all →- SevereMicro Bird Recalls Transit Buses Due to Fire Risk from Loose Power Cable
NHTSA · 2025-12-05
- High
- SevereHeartMate Mobile Power Unit electrical malfunction may cause loss of power
FDA (Devices) · 2025-04-23
- Severe
- Severe
Similar recalls
- HighMagstim Recalls Neurosign V4 Intraoperative Nerve Monitors Due to Cable Damage Risk
FDA (Devices) · 2020-09-02
- HighStryker C2 Nerve Monitor Recalled for Potential Nerve Injury Risk
FDA (Devices) · 2019-10-09
- HighStryker Recalls C2 Nerve Monitor Due to Usability Design Issues
FDA (Devices) · 2019-10-09
- HighMagstim Recalls Rapid2 Stim Interconnecting Cables Due to Potential Failure
FDA (Devices) · 2020-05-20
- ModerateNuvasive NVM5 EMG Module Recalled for Incorrect Component
FDA (Devices) · 2020-07-29
- ModerateNeurovision Nerveana x5 RLN Monitoring Kit Recalled for Insulation Defect
FDA (Devices) · 2020-09-02