GE Healthcare Recalls Centricity Enterprise Archive Software Due to Data Loss Risk
GE Healthcare is recalling Centricity Enterprise Archive software versions 4.0.x due to a potential data loss issue associated with HL-7 updates.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard is a potential data loss issue, which is a significant operational risk but does not involve direct physical injury or death, fitting the criteria for a moderate severity score.
Plain-English summary
GE Healthcare is recalling 115 units of Centricity Enterprise Archive software, versions 4.0.x. This software is used for receiving, archiving, and sending medical data and communicates with DICOM devices using the DICOM protocol. The recall is due to a potential data loss issue that can occur when the cache disk is almost full and HL-7 updates result in empty path names.
The affected software was distributed nationwide, including in Arizona, Ohio, and Washington. The intended users for Centricity Enterprise Archive are service personnel, system administrators, and non-clinical users. As of version 4.0, XDS-enabled systems can also communicate with the archive using XDS and XDS-1 profiles.
The source text does not provide specific instructions for consumers or affected users on how to proceed with the recall or remediation.
The recalled product
- Product
- Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction…
- Manufacturer
- Ge Healthcare It
- Category
- Medical Device — Software
- Hazard
- Affected units
- 115
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Centricity Enterprise Archive versions: 4.0.x
Distribution
Related recalls
Same manufacturer · Ge Healthcare It
See all →- ModerateGE Healthcare Recalls Centricity Enterprise Archive Software Due to Data Loss Risk
FDA (Devices) · 2016-09-14
- HighGE Healthcare Recalls Centricity Laboratory Software Due to Data Deletion Errors
FDA (Devices) · 2016-09-07
- ModerateGE Centricity PACS Workstation Recalled for Image Grouping Errors
FDA (Devices) · 2016-08-24
- HighGe Healthcare Recalls Imagecast PACS Software Due to Image Context Defect
FDA (Devices) · 2016-08-03
- LowGE Centricity Laboratory System Software Recall for Calculation Errors
FDA (Devices) · 2014-02-05
Same hazard · data loss
See all →- HighPhilips Fluoroscopy System Recall Over Hard Disk Drive Issue
FDA (Devices) · 2026-08-12
- SeverePhilips Allura X-Ray Systems Recalled Over Unauthorized Hard Disk Drives
FDA (Devices) · 2026-08-12
- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighMint Lesion Software Data Loss During Server Connection Interruption
FDA (Devices) · 2026-05-13
Similar recalls
- ModerateGE Healthcare Recalls Centricity Enterprise Archive Software Due to Data Loss Risk
FDA (Devices) · 2016-09-14
- HighGE Healthcare Recalls Centricity Enterprise Archive Software Due to Data Association Errors
FDA (Devices) · 2026-09-23
- HighGE Centricity Web Software Recalled for Patient Safety Concerns
FDA (Devices) · 2013-12-11
- ModerateGE Healthcare Recalls Centricity Laboratory System Software Due to Data Loss
FDA (Devices) · 2013-07-03
- HighGE Healthcare Recalls Centricity PACS Foundation Software
FDA (Devices) · 2019-07-10
- ModerateGE Centricity Laboratory System Recall for Free-Text Data Loss
FDA (Devices) · 2014-01-08