Fresenius NaturaLyte Liquid Acid Concentrate Recall for Low Sodium Levels
Fresenius Medical Care is recalling NaturaLyte Liquid Acid concentrate bottles because sodium and/or calcium levels were below manufacturing specifications.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk medical device used in hemodialysis where chemical levels are below specifications, representing a risk of harm where injury has not yet been reported.
Plain-English summary
Fresenius Medical Care Holdings, Inc. is recalling NaturaLyte Liquid Acid concentrate bottles, Model Number 08-4231-0. The recall involves 1,726 cases distributed nationwide in the United States. The affected lot numbers are 20LXAC052 and 20LXAC088.
The recall was initiated because the sodium and/or calcium levels in these concentrates were below the manufacturing specifications. These products are used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Healthcare facilities and consumers should stop using the affected products and contact the recalling firm for further instructions.
The recalled product
- Product
- NaturaLyte Liquid Acid concentrate bottles, Model Number 08-4231-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Affected units
- 1,726
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot Numbers: 20LXAC052 Additional affected lot as of 3/17/21: 20LXAC088
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 19 recalled by Fresenius Medical Care Holdings, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 19 products in this recall →Related recalls
Same manufacturer · Fresenius Medical Care Holdings, Inc.
See all →- Moderate5008X Blood Tubing Set connection errors FDA Class II recall
FDA (Devices) · 2026-07-15
- HighBicarby dialysate bags carry an incorrect GTIN on the label
FDA (Devices) · 2026-06-17
- HighBicarby dialysate luer-lock connections may leak fluid onto floors
FDA (Devices) · 2026-06-17
- SevereDELFLEX Peritoneal Dialysis Solution Recall Due to Potential Sterility Issues
FDA (Drugs) · 2026-05-13
- HighDELFLEX Peritoneal Dialysis Solution bags recalled for sterility risk
FDA (Drugs) · 2026-05-13
Same hazard · chemical mislabeling
See all →- ModerateDialysis concentrate labeling error for chemical composition
FDA (Devices) · 2022-11-16
- HighFresenius NaturaLyte Liquid Acid Concentrate Recall for Low Sodium and Calcium Levels
FDA (Devices) · 2021-01-20
- High
Similar recalls
- ModerateFresenius NaturaLyte Liquid Acid Concentrate Recalled for Potassium Concentration Error
FDA (Devices) · 2018-03-21
- HighFresenius Naturalyte Liquid Acid Concentrate Recall for Conductivity Deviation
FDA (Devices) · 2021-01-20
- SevereFresenius Naturalyte Liquid Acid Concentrate Recalled for Conductivity Deviation
FDA (Devices) · 2021-01-20
- HighFresenius Naturalyte Liquid Acid concentrate bottles recalled for conductivity deviation
FDA (Devices) · 2021-01-20
- ModerateFresenius NaturaLyte Liquid Acid Concentrate Recalled for Mislabeled Part Numbers
FDA (Devices) · 2018-06-13
- HighFresenius Naturalyte Liquid Acid concentrate drums recalled for conductivity deviation
FDA (Devices) · 2021-01-20