HeartWare HVAD System: Updated Cleaning Instructions for Power Connectors
Updated cleaning instructions are being provided for the HeartWare HVAD System's power connectors. Improper cleaning can remove protective lubricant required for proper power switching.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This Class II recall provides updated cleaning instructions to prevent improper maintenance that could affect power switching. No illnesses, injuries, or deaths are reported in the source, making this a precautionary instructional update.
Plain-English summary
The HeartWare Ventricular Assist Device (HVAD) System is a mechanical cardiac pump used to support heart function in patients with advanced heart failure. Heartware, Inc. is providing updated cleaning instructions for the Controller AC Adapter, DC Adapter, and Battery connectors.
The update clarifies that the power source connector pins should not be cleaned, as doing so can remove a protective lubricant applied to the pins. This lubricant is essential for proper power switching functionality.
The recall affects approximately 197,889 units distributed throughout the United States and more than 50 countries worldwide. Patients and healthcare providers who use or service these devices should review and follow the updated cleaning instructions provided by Heartware, Inc. to ensure proper device operation and maintenance.
The recalled product
- Product
- HeartWare Ventricular Assist Device (HVAD) System
- Manufacturer
- Heartware, Inc.
- Hazard
- connector-damage
- power-failure
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- All lots/serials for the following product model numbers: 1440
- 1400
- 1401
- 1401AU
- 1401BR
- 1401CA
- 1401CH
- 1401DE
- 1401GB
- 1401IL
- 1401IN
- 1401IT
- 1401JP
- 1401US
- 1403US
- 1407AR
- 1407AU
- 1407BR
- 1407CA
- 1407CA-CLIN
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- SevereMEDLINE Medical Procedure Kits Containing Bupivacaine Recalled Due to Drug Ineffectiveness
FDA (Devices) · 2026-07-29
- HighSafeOp Patient Interface Neuromonitoring Device Recalled Due to Software Issue
FDA (Devices) · 2026-07-29
- HighPenumbra System RED 68 Reperfusion Catheter Recall Due to Fracture Risk
FDA (Devices) · 2026-07-29
- ModerateFARASTAR Ablation Generator EU Medical Device Recall for Labeling Discrepancy
FDA (Devices) · 2026-07-29
- High