Siemens SOMATOM Go.Up CT Table Recall for Potential Instability
Siemens Medical Solutions is recalling six SOMATOM Go.Up CT system tables because improperly tightened bolts may loosen, causing the table to become unstable.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a structural defect in a medical device that could lead to instability and potential injury, meeting the criteria for a Severe rating (structural defect/significant injury risk).
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling six units of the SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number 11061628. The recall involves specific serial numbers: 108054, 108059, 108062, 108069, 108076, and 108077. These devices were distributed in the United States in the states of Florida, Missouri, New York, and Tennessee.
The recall is being issued because the nuts on the bolts of the table may not have been adequately tightened. This defect has the potential to allow the bolts to loosen, with several bolts potentially loosening at the same time. If this occurs, the table may become unstable.
Healthcare facilities that possess these specific serial numbers should contact Siemens Medical Solutions USA, Inc. for instructions on how to resolve the issue. The source text does not report any injuries or illnesses associated with this recall.
The recalled product
- Product
- SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a…
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 6
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (6)
- Serial Numbers: 108054
- 108059
- 108062
- 108069
- 108076
- 108077
Distribution
Part of a larger recall action
This product is one of 2 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · structural defect
See all →- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- SevereVolvo Recalls 2025-2026 VNL Trucks for Loose Structural Bolts
NHTSA · 2026-10-15
- Severe
- SevereVolvo Recalls EX30 and EX30CC Vehicles for Head Restraint Defect
NHTSA · 2026-10-09
- Severe
Similar recalls
- HighSiemens SOMATOM go.Up CT Systems Recalled for Falling Injector Holder
FDA (Devices) · 2019-09-04
- ModerateSiemens SOMATOM go.Up CT Systems Recalled for Performance Issues
FDA (Devices) · 2019-07-10
- HighSiemens SOMATOM go.Top CT Scanner Recall for Falling Injector Holder
FDA (Devices) · 2019-09-04
- ModerateSiemens SOMATOM go.Up CT Scanner Recalled for Performance Issues
FDA (Devices) · 2019-07-10
- ModerateSiemens SOMATOM Go.Up CT Scanner Recalled for Software Issue
FDA (Devices) · 2018-06-06
- HighSiemens SOMATOM go.Now CT Systems Recalled for Injector Holder Risk
FDA (Devices) · 2019-09-04