Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0649-2020·Announced 2019-12-18

Integra CUSA Clarity System Recall for Sterilization Tray Drying Defect

Integra LifeSciences is recalling the C7000 CUSA Clarity System because the 36 kHz Sterilization Tray does not meet the 30-minute drying time criteria stated in the manual.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: This is a Class II recall involving a labeling/documentation discrepancy regarding drying times. No injuries or illnesses are reported, and the hazard is theoretical, fitting the criteria for a moderate severity score.

Plain-English summary

Integra LifeSciences Corp. is recalling the C7000 - CUSA Clarity Ultrasonic Tissue Ablation System. The recall involves the 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system.

The recall is due to a defect where the system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components, but 40 minutes is required for complete drying.

The affected products were distributed in the United States and internationally. Users should contact the manufacturer for instructions on how to proceed.

The recalled product

Product
C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Croatian OPS Manual P/N 60905786

Distribution

Part of a larger recall action

This product is one of 42 recalled by Integra LifeSciences Corp. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 42 products in this recall →