ModerateFDA (Devices)·Z-0594-2019·Announced 2018-12-19

Helena Laboratories Recalls Cascade Abrazo c-ACT-LR Test Cards

Helena Laboratories is recalling Cascade Abrazo c-ACT-LR test cards due to waveform variations that may result in inaccurate clotting time measurements.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves potential diagnostic inaccuracy due to waveform variations, but no specific injuries, illnesses, or deaths are reported in the source text.

Plain-English summary

Helena Laboratories, Inc. is recalling 12 kits of Cascade Abrazo c-ACT-LR Test Cards, Lot 1-15-5701, Model/Catalog Number 5701. These in vitro diagnostic test cards are intended to monitor the effect of low to moderate heparin on coagulation in fresh whole blood when used with the Cascade Abrazo analyzer.

The recall was initiated because some cards in this lot experienced variations in wave form, resulting in some cards with shorter or longer times than expected. This defect may lead to inaccurate activated clotting time results.

The affected products were distributed in the United States to the state of Texas and internationally to Australia, the Netherlands, Italy, and England. Users of these test cards should stop using the affected lot and contact Helena Laboratories for further instructions or a replacement.

The recalled product

Product
Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) FOR EXPORT ONLY The Cascade Abrazo c-ACT-LR test cards are activated clotting time tests to be used with the Cascade Abrazo analyzer and are intended
Affected units
12

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot Codes: 1-15-5701 Model/Catalog Number: 5701

Distribution

Distributed in 1 state: