Bronchoscopes recalled after leak testing with equipment outside its operating range
Auris Health is recalling 1,477 MONARCH Bronchoscopes because the product may have been leak tested using equipment operating outside its expected range.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: Scored 2 because the record cites a leak testing process deviation rather than a confirmed device defect or reported harm.
Plain-English summary
The MONARCH Bronchoscope, model MBR-000211-B, from Auris Health, is being recalled.
The firm describes the potential that product was leak tested with equipment outside of its expected operating range, stating in the record that this resulted in bronchoscopes that may not be susceptible to leaks.
The recall covers 1,477 units, identified by serial number. The product was distributed nationwide, including Alabama, Arizona, California, Colorado, Connecticut, the District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kentucky, Massachusetts, Michigan, Mississippi, Montana, North Carolina, Nebraska, New Jersey, New Mexico and New York. The FDA has classified the action as Class II.
The enforcement record describes a testing process deviation rather than a confirmed device defect, and reports no injuries. Facilities holding affected units should follow the firm's instructions.
The recalled product
- Product
- MONARCH Bronchoscope. Model Number: MBR-000211-B
- Manufacturer
- Auris Health, Inc
- Category
- Medical Device — Bronchoscope
- Affected units
- 1,477
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1311678 (11/14/2024)
- 1311679 (11/14/2024)
- 1311680 (11/14/2024)
- 1311681 (11/14/2024)
- 1311682 (11/14/2024)
- 1311683 (11/14/2024)
- 1311684 (11/14/2024)
- 1311685 (11/14/2024)
- 1311686 (11/14/2024)
- 1311687 (11/14/2024)
- 1311688 (11/14/2024)
- 1311722 (11/15/2024)
- 1311723 (11/15/2024)
- 1311724 (11/15/2024)
- 1311725 (11/15/2024)
- 1311726 (11/15/2024)
- 1311727 (11/15/2024)
- 1311730 (11/15/2024)
- 1311731 (11/15/2024)
- 2404220489 (11/21/2024)
Distribution
Related recalls
Same manufacturer · Auris Health, Inc
See all →- HighBronchoscopy robots recalled over arms that may move toward patient-left
FDA (Devices) · 2026-02-25
- HighMonarch Platform Urology Device Recalled for Power Cord Electrical Hazard
FDA (Devices) · 2024-10-16
- HighEndoscopy and Bronchoscopy Systems Power Cord Defect Risk
FDA (Devices) · 2024-10-16
- HighMonarch Platform Bronchoscope Software Issue Causes Flipped Display
FDA (Devices) · 2023-12-13
- HighMonarch Platform Bronchoscopy Device Recalled for Software Image Inversion
FDA (Devices) · 2023-12-13