Zimmer Recalls Persona Knee Implants to Reduce Tibial Perforation Risk
Zimmer, Inc. is recalling 318 units of Persona Knee System tibial implants and tools to reduce the risk of tibial perforation and fracture.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II and involves a structural defect in a medical device that poses a risk of significant injury (tibial bone fracture).
Plain-English summary
Zimmer, Inc. is recalling 318 units of the Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile implants, along with associated tools including the Tibial Drill Guide, Broach, and Sizing Plate. The affected products include Size C Left and Right tibial components and specific lots of the Persona Cemented Tibial Drill.
The recall is being conducted to reduce the risk of tibial perforation, which could result in tibial bone fracture due to the stress riser created by a perforation, or tibial plate loosening. The affected units were distributed nationwide in the United States and internationally to Germany and Europe.
Healthcare providers and patients who have received these specific implants or tools should contact Zimmer, Inc. for information regarding the replacement with enhanced products and instruments.
The recalled product
- Product
- Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile SIZE C, LEFT; SIZE C RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis…
- Manufacturer
- Zimmer, Inc.
- Affected units
- 318
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Part Number 42-5320-064-01 TIBIA SIZE C
- LEFT Lots: 62018336
- 62018337
- 62033380
- 62051020
- 62060739
- 62068413
- 62076438
- 62079740
- 62091360
- 62117665
- and 77001708
- Part Number 42-5320-064-02 TIBIA SIZE C RIGHT Lot: 62018338
- 62018339
- 62033383
- 62051021
- 62060740
- 62060742
- 62068415
- 62076439
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 17 recalled by Zimmer, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 17 products in this recall →Related recalls
Same manufacturer · Zimmer, Inc.
See all →- HighZimmer Off-Axis Alliance Glenoid Reamer Guide Plastic Breakage Recall
FDA (Devices) · 2026-07-01
- HighZimmer Off-Axis Alliance Glenoid Reamer Guide recalled for breakage
FDA (Devices) · 2026-07-01
- HighZimmer Shoulder Reamer Guide System Recalled for Plastic Breakage
FDA (Devices) · 2026-07-01
- ModeratePersona Revision Trabecular Metal Femoral Distal Augment recalled for packaging defect
FDA (Devices) · 2026-07-01
- HighZimmer Off-Axis Comprehensive Shoulder System Reamer Guide Recall
FDA (Devices) · 2026-07-01
Same hazard · bone fracture
See all →- HighFathom Pedicle Retractor Blades Recalled for Manufacturing Defect
FDA (Devices) · 2024-07-10
- HighKnee Patella Drills Recalled for Incorrect Diameter Markings
FDA (Devices) · 2023-04-19
- HighShoulder Surgical Navigation System Warning for Bone Stability
FDA (Devices) · 2022-12-14
- HighSmith & Nephew Recalls Journey II XR Tibial Posterior Keel Punch
FDA (Devices) · 2019-04-03
- SevereSmith & Nephew Recalls Journey II XR Tibial Posterior Keel Punch
FDA (Devices) · 2019-04-03
Similar recalls
- SevereZimmer Recalls Cemented Tibial Drills Due to Bone Perforation Reports
FDA (Devices) · 2012-10-10
- ModerateZimmer Persona Stemmed 5 Degree Cemented Tibia Recalled for Cleaning Failure
FDA (Devices) · 2014-11-19
- HighZimmer Recalls Persona Knee Instrument Trays Due to Sterilization Risk
FDA (Devices) · 2014-08-27
- SevereZimmer Persona Knee Trial Articular Surface Provisional Breakage Recall
FDA (Devices) · 2014-09-10
- HighZimmer Recalls Persona Knee Trial Articular Surface Provisionals Due to Breakage
FDA (Devices) · 2014-09-10
- ModerateZimmer Biomet Recalls Persona Knee System Components Due to Missing Parts
FDA (Devices) · 2018-04-04