Siemens Artis Fluoroscopic X-ray Systems Recalled for Display Soldering Defect
Siemens is recalling 17 Artis fluoroscopic X-ray systems due to improper soldering in display modules that could cause video signal loss.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source explicitly states the issue has not been observed in the field and only sporadic cases were seen in factory screening, fitting the criteria for a voluntary precautionary recall with no reported harm.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 17 Artis interventional fluoroscopic X-ray systems with large displays. The recall is due to improper soldering in a limited number of large display bypass module components from a specific production lot.
This defect could potentially cause a loss of video signal. The source text notes that this issue has not been observed in the field and only sporadic cases have been observed in factory screening.
The affected units were distributed to hospitals and facilities in Illinois, Wisconsin, Massachusetts, Florida, Alabama, Tennessee, North Dakota, Ohio, Texas, California, New Jersey, and Michigan. Affected model numbers include 10094135, 10094137, 10094141, 10848280, 10848281, 10848282, and 10848355. Affected serial numbers include 123026, 147909, 123031, 103030, 147912, 147910, 147921, 147919, 131005, 121096, 125004, 137551, 121088, 109104, 109105, 109107, and 121085.
The recalled product
- Product
- Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Hazard
- Affected units
- 17
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model numbers 10094135
- 10094137
- 10094141
- 10848280
- 10848281
- 10848282
- 10848355. Serial Numbers 123026
- 147909
- 123031
- 103030
- 147912
- 147910
- 147921
- 147919
- 131005
- 121096
- 125004
- 137551
- 121088
- 109104
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · malfunction
See all →- HighFresenius BlueStar Hemodialysis Machines Recalled for Alarm Defects
FDA (Devices) · 2026-09-02
- ModerateOlympus 4K UHD LCD Monitors Recalled Due to Potential Connector Faults
FDA (Devices) · 2026-07-29
- High
- ModeratePhilips Azurion 3M15 Medical Imaging Systems Recalled Due to Software Issue
FDA (Devices) · 2026-06-10
- SevereElectric Start Pressure Washers Recalled for Carbon Monoxide Hazard
CPSC · 2026-05-14