Natus Olympic Cool-Cap Recalled Due to Frozen Screen Malfunction
Natus Medical is recalling 324 Olympic Cool-Cap systems because the control module screen may freeze, causing the device to stop cooling infants without visible alarm.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "The Olympic Cool-Cap's Control Module has experienced a frozen screen during use. When this occurs, the on-screen information remains on display, but the system is no longer providing cooling treatment to the infants. The visible clock…"
Plain-English summary
Natus Medical Incorporated is recalling 324 units of the Olympic Cool-Cap System, a selective head cooling device used to treat hypoxic-ischemic encephalopathy (HIE) in infants. The recall is due to a defect in the Control Module where the screen may freeze during use. When this occurs, the on-screen information remains visible, but the system stops providing cooling treatment to the infant. The visible clock in the upper right corner of the screen stops advancing, which may not be immediately obvious to users.
The affected devices were distributed worldwide, including nationwide in the United States and in numerous international countries. The specific part number for the recalled product is 60010.
Healthcare facilities should stop using the affected devices and contact Natus Medical Incorporated for instructions on repair or replacement. Consumers should not attempt to repair the device themselves.
The recalled product
- Product
- The product has the Catalogue/Part number 60010. Olympic Cool-Cap System is a selective head cooling system for treatment of hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers.…
- Manufacturer
- Natus Medical Incorporated
- Affected units
- 324
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1008
- 1009
- 1011
- 1012
- 1014
- 1016
- 1017
- 1018
- 1020
- 1021
- 1022
- 1023
- 1024
- 1025
- 1027
- 1028
- 1029
- 1030
- 1031
- 1032
Distribution
Related recalls
Same manufacturer · Natus Medical Incorporated
See all →- HighExternal Drainage System Collection Bags Recalled for Sterility Assurance Issues
FDA (Devices) · 2024-01-10
- HighEDS 3 CSF External Drainage System recalled for incomplete sterility testing
FDA (Devices) · 2024-01-10
- HighNatus Cranial Access Kit Recalled Due to Defective Sterile Surgical Drapes
FDA (Devices) · 2022-12-28
- ModerateNatus Medical Recalls Olympic Pasteurmatic 3000 and 3500 Disinfection Systems
FDA (Devices) · 2017-07-19
- HighNatus Medical Recalls neoBLUE 2 Phototherapy Lights for Intensity Defect
FDA (Devices) · 2017-05-10
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- HighNatus Olympic Cool-Cap System Recalled Due to Power Supply Failure
FDA (Devices) · 2012-06-27
- ModerateNatus Olympic Brainz Monitor Recalled for Missing Functional Tests
FDA (Devices) · 2020-04-08
- HighOlympus UES-40 Electrosurgical Unit Recalled for Overheating and Smoke
FDA (Devices) · 2013-05-15
- HighXLTEK EMU40EX EEG Headbox Recalled for Component Degradation
FDA (Devices) · 2018-08-08
- ModerateNatus Medical Recalls Olympic Pasteurmatic 3000 and 3500 Disinfection Systems
FDA (Devices) · 2017-07-19
- ModerateNatus neoBLUE Phototherapy Systems Recalled for Fiberoptic Pad Failure
FDA (Devices) · 2016-12-21