Agency classClass IIOur severity:SevereFDA (Devices)·Z-0437-2027·Announced 2026-11-18

Tandem Mobi Insulin Pump Software Defect Recalled by FDA

Tandem Diabetes Care is recalling the Mobi Insulin Pump due to a software defect that may cause under-delivery of insulin and hyperglycemia.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II, and the source describes a software defect that may result in unrecognized under-delivery of insulin and hyperglycemia. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Tandem Diabetes Care, Inc. is recalling the Tandem Mobi Insulin Pump with Interoperable Technology (Model/Catalog Number: 1015472) in the United States and the Czech Republic. The recall involves approximately 4,086 users who are using the Android Tandem Mobi Mobile app with software versions 3.5.4, 3.5.2.2, or 3.5.2.1.

The recall is due to a software defect in the mobile application. When the entered glucose value equals the user's active target glucose value, the immediate (Now) portion of an Extended Bolus is incorrectly reduced to 0 units. While the deferred (Later) portion is delivered, the total insulin delivered may be less than intended. The application does not provide a distinct alert that the immediate portion was omitted, which may result in unrecognized under-delivery of insulin and hyperglycemia.

Consumers should contact Tandem Diabetes Care for instructions on how to address the software defect and ensure safe insulin delivery.

The recalled product

Product
Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472
Affected units
4,086

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4
  • 3.5.2.2
  • and 3.5.2.1

Distribution

Distributed nationwide across the United States.