bkSpecto ultrasound systems recalled over battery guidance gaps in manuals
B-K Medical is recalling 1,697 Ultrasound System 1300 bkSpecto units because the user guide and service manual do not adequately state battery replacement intervals, end-of-life handling or safety precautions.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: Scored 2 as a labeling deficiency: the record cites missing instructions rather than a device defect, and states no consequence or reported harm.
Plain-English summary
The Ultrasound System 1300, bkSpecto with battery, models 1300-21 and 1300-S1, is being recalled.
GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
The recall covers 1,697 units, including GTIN 05704916001285, distributed worldwide including nationwide in the United States and to countries including Algeria, Argentina, Australia and Austria. The FDA has classified the action as Class II.
The enforcement record describes the documentation gap without stating a health consequence, and reports no injuries. Sites operating an affected system should follow the firm's instructions.
The recalled product
- Product
- Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, Ultrasonic
- Manufacturer
- B-K Medical A/S
- Affected units
- 1,697
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1) 1300-21: GTIN 05704916001285 and pending
- Serial Numbers: 2002690
- 2008466
- 2008495
- 2006774
- 5012683
- 5013881
- 5012684
- 2008725
- 2008280
- 2004419
- 5012980
- 2007101
- 2009029
- 5014333
- 2003852
- 2002851
- 2002142
- 2005860
- 5013937
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by B-K Medical A/S under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · B-K Medical A/S
See all →- Moderatebk3000 ultrasound systems recalled over incomplete battery replacement guidance
FDA (Devices) · 2025-11-12
- ModerateUltrasound systems recalled over battery guidance missing from the manuals
FDA (Devices) · 2025-11-12
- Moderatebk3500 ultrasound systems recalled over incomplete battery safety instructions
FDA (Devices) · 2025-11-12
- ModerateFlex Focus ultrasound carts recalled over incomplete battery safety instructions
FDA (Devices) · 2025-11-12
- ModeratebkActiv ultrasound systems recalled over battery guidance missing from manuals
FDA (Devices) · 2025-11-12
Same hazard · incomplete instructions
See all →- HighECG cable shipped with incomplete cleaning and disinfection instructions
FDA (Devices) · 2025-02-12
- ModerateTENS Devices Recalled for Missing Instruction Manual Warnings
FDA (Devices) · 2023-12-13
Similar recalls
- HighUltrasound Systems Recalled for Loose Power Connections Causing Thermal Hazard
FDA (Devices) · 2024-10-30
- HighGE Vivid i Ultrasound Systems Battery Failure Recall
FDA (Devices) · 2022-12-28
- HighBecton Dickinson Recalls Site~Rite 8 Ultrasound Systems Due to Battery Indicator Defect
FDA (Devices) · 2019-12-04
- HighUltrasound system recalled due to measurement error in Dual Live Compare feature
FDA (Devices) · 2024-01-31
- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
- HighGE Vivid q Ultrasound Systems: Battery Fire Risk Recall
FDA (Devices) · 2022-12-28