The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0380-2027·Announced 2026-11-18

TransMedics Recalls OCS Heart Perfusion Sets Due to Startup Error

TransMedics, Inc. is recalling OCS Heart Perfusion Sets because an unprogrammed Front End Board may prevent the system from starting.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard is a functional defect where an unprogrammed Front End Board prevents the system from starting, displaying an error message. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

TransMedics, Inc. is recalling 15 OCS Heart Perfusion Sets, Model/Catalog Number 1200, Software Version rev 3. The recall involves the OCS Heart Perfusion Module used for organ transport. The affected units were distributed in the United States (CA, FL, IL, MA, MN, NC, NE, NY, TX) and Serbia.

The recall is due to a potential defect where the Perfusion Module may have an unprogrammed Front End Board (FEB). This issue prevents the system from proceeding beyond the initial startup process. The OCS system displays an error message stating "Unsupported Perfusion Module Present" and prevents further system use.

The source text does not report any illnesses, injuries, or deaths associated with this recall. Users should check their equipment against the specific serial numbers listed in the official recall notice and contact TransMedics, Inc. for further instructions.

The recalled product

Product
OCS Heart Perfusion Set Model/Catalog Number: 1200 Software Version: rev 3 Product Description: OCS Heart Perfusion Module, included in the Heart Perfusion Set used for organ transport.
Manufacturer
TransMedics, Inc.
Affected units
6

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • 1. Perfusion Module SN: 4822-8-22505
  • Description: OCS Heart Perfusion Set
  • EXP Date: 4/9/29
  • UDI: *+B00948225/$$+74822-8-22505/14D20290409C*
  • 2. Perfusion Module SN: 4822-8-23601
  • EXP Date: 7/27/29
  • UDI: *+B00948225/$$+74822-8-23601/14D20290727D*
  • 3. Perfusion Module SN: 4822-8-22496
  • UDI: *+B00948225/$$+74822-8-22496/14D20290409L*
  • 4. Perfusion Module SN: 4822-8-23497
  • UDI: *+B00948225/$$+74822-8-23497/14D20290409N*
  • 5. Perfusion Module SN: 4822-8-22479
  • UDI: *+B00948225/$$+74822-8-22479/14D20290409M*
  • 6. Perfusion Module SN: 4822-8-22504
  • UDI: *+B00948225/$$+74822-8-22504/14D20290409B*
  • 7. Perfusion Module SN: 4822-8-23604
  • EXP Date: 7/13/29
  • UDI: *+B00948225/$$+74822-8-23604/14D20290713B*
  • 8. Perfusion Module SN: 4822-8-23600
  • UDI: *+B00948225/$$+74822-8-23600/14D202907137*

Distribution

Distributed nationwide across the United States.