The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0347-2027·Announced 2026-11-11

Shimadzu X-Ray System C-Arm Bolts May Not Be Torqued to Specification

Shimadzu is recalling 118 interventional fluoroscopic X-ray systems because C-arm mounting bolts may not be torqued to specification, potentially compromising structural stability.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source identifies a structural defect where mounting bolts may not be torqued to specification, potentially compromising structural stability. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Shimadzu Corporation Medical Systems Division is recalling 118 units of the Trinias Opera Smart, Opera Grand, and Smart OBL Interventional Fluoroscopic X-Ray Systems. The recall involves specific C-arm support models (MH-500, MH-60) distributed in South Carolina, Texas, Florida, Tennessee, California, and New York.

The recall is being conducted because the mounting bolts securing the C-arm drive unit may not be torqued to specification. This defect potentially compromises the structural stability and safe operation of the X-ray system.

The source text does not report any illnesses, injuries, or deaths associated with this recall. Healthcare facilities should contact Shimadzu Corporation Medical Systems Division for instructions on how to address the defect.

The recalled product

Product
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
Affected units
118

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (13)

  • M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241
  • M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241
  • 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240
  • M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240
  • M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240
  • 41F0F40E5001/ N/A/ (01)04540217068937(11)240
  • 41F0F41E5001/ N/A/ (01)04540217068944(11)240
  • 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240
  • M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240
  • 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240
  • 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231
  • 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231
  • M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230

Distribution

Distributed nationwide across the United States.

The notice also names 6 states specifically:

Same manufacturer · SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION

See all →