The Recall Desk
HighFDA (Devices)·Z-0333-2026·Announced 2025-11-05

Chemotherapy administration sets recalled because port weld may separate and leak

ICU Medical is recalling 7,950 20-drop administration sets with an integrated ChemoLock port because the port weld of the drug transfer device may separate or break during use and leak.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: a weld that can break and leak on a drug transfer device is a hazard, and no injuries are reported.

Plain-English summary

The ICU Medical 30 inch (76 cm) approximately 3.6 mL, 20 drop administration set with integrated ChemoLock port drip chamber, ChemoLock with red cap and bag hanger, REF CL3511, is being recalled.

The port weld of the drug transfer device may separate or break during use and potentially result in a leak.

The recall covers 7,950 units carrying UDI 10887709057026, from lot 14400616. Distribution was worldwide, including nationwide in the United States. The FDA has classified the action as Class II.

The enforcement record describes the weld failure and reports no injuries. Because these sets are used to transfer hazardous drugs, facilities should quarantine the affected lot and follow ICU Medical's instructions.

The recalled product

Product
ICU Medical 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, ChemoLock w/Red Cap, Bag Hanger REF: CL3511
Manufacturer
ICU Medical, Inc.
Affected units
7,950

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI:10887709057026/Lot: 14400616

Distribution

Distributed nationwide across the United States.

The notice also names 38 states specifically:

Part of a larger recall action

This product is one of 38 recalled by ICU Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 38 products in this recall →