The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0325-2027·Announced 2026-11-11

Artivion Recalls CryoPatch SynerGraft Pulmonary Human Cardiac Patch

Artivion, Inc. is recalling a CryoPatch SynerGraft Pulmonary Human Cardiac Patch because the donor was older than the specified age limit, raising concerns about potential graft dysfunction.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source identifies a potential hazard of premature graft dysfunction due to the donor age being outside specifications. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Artivion, Inc. is recalling one unit of the CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model SGP010. The recall is due to the cryopreserved pulmonary human cardiac patch being taken from a 64-year-old human donor, which is outside the specified age range of newborn to 60 years for cardiac tissue donors.

The patch was shipped on May 22, 2024, and implanted on June 5, 2024. The distribution was nationwide in the state of Illinois. The specific unit is identified by Serial/Donor Number 12617554/177871 and UDI-DI 00877234000478.

Due to limitations in long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded. The source text does not report any illnesses, injuries, or deaths associated with this recall.

The recalled product

Product
CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010
Manufacturer
Artivion, Inc
Affected units
1

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (4)

  • Serial/Donor Number: 12617554/177871
  • UDI-DI: 00877234000478
  • Shipped: 22May2024
  • Implanted: 05June2024

Distribution

Distributed nationwide across the United States.