Philips Azurion 7 B20 Recalled for Unspecified Solid-State Drive Reliability Issues
Philips is recalling 457 Azurion 7 B20 systems because unspecified solid-state drives may cause loss of imaging functionality, blank displays, or inability to store or access images.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall involves a Class II medical device where the hazard is reduced reliability leading to loss of functionality or data access, but the source does not report any illnesses, injuries, or deaths.
Plain-English summary
Philips Medical Systems Nederland B.V. is recalling 457 units of the Philips Azurion 7 B20 (Model Numbers 722068 and 722226). The recall is due to the potential installation of solid-state drives (SSDs) that are not specified in the Medical Device File. These non-specified SSDs may have different performance characteristics, including reduced reliability over time.
The source states that these reliability issues may result in the loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images. The affected units were distributed worldwide, including in the United States and numerous other countries.
The source does not provide specific instructions for consumers or healthcare facilities on how to proceed with the recall, nor does it report any illnesses, injuries, or deaths associated with this issue.
The recalled product
- Product
- Philips Azurion 7 B20. Model Numbers: 722068, 722226.
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Category
- Medical Device — Imaging System
- Affected units
- 457
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1. Model Number: 722068. UDI Numbers: (01)00884838085367(21)178
- (01)00884838085367(21)147
- (01)00884838085367(21)205
- (01)00884838085367(21)170
- (01)00884838085367(21)494
- (01)00884838085367(21)247
- (01)00884838085367(21)149
- (01)00884838085367(21)153
- (01)00884838085367(21)85
- (01)00884838085367(21)320
- (01)00884838085367(21)89
- (01)00884838085367(21)172
- (01)00884838085367(21)505
- (01)00884838085367(21)271
- (01)00884838085367(21)443
- (01)00884838085367(21)354
- (01)00884838085367(21)196
- (01)00884838085367(21)73
- (01)00884838085367(21)285
- (01)00884838085367(21)406
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
See all →- HighPhilips Azurion 7 M12 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 7 M20 Medical Device Recall for Unspecified SSDs
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 5 M12 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 5 M20 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 3 M12 Recalled for Unspecified Solid-State Drive Reliability Issues
FDA (Devices) · 2026-11-04
Similar recalls
- ModeratePhilips Azurion 7 M20 Medical Device Recall for Unspecified SSDs
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 7 B12 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 5 M20 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- HighPhilips Azurion 7 M12 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 3 M12 Recalled for Unspecified Solid-State Drive Reliability Issues
FDA (Devices) · 2026-11-04
- ModeratePhilips Azurion 5 M12 Recalled for Unspecified Solid-State Drives
FDA (Devices) · 2026-11-04