The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0245-2027·Announced 2026-10-28

Boston Scientific Recalls Defective Direxion Torqueable Microcatheters

Boston Scientific is recalling 382 Direxion Torqueable Microcatheters due to insufficient UV curing of the hydrophilic coating and raised ridges on nitinol sleeves.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II due to manufacturing defects involving insufficient UV curing and laser cutting misalignment. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Boston Scientific Cork, Ltd. is recalling 382 units of the Direxion Torqueable Microcatheter, Material Number M001195470. The affected devices were distributed worldwide, including the United States, Puerto Rico, and numerous international markets.

The recall is due to a manufacturing defect involving insufficient UV curing of the hydrophilic coating on the outer surface of the distal segment of the microcatheters. Additionally, the devices have raised ridges on the nitinol sleeves caused by laser cutting misalignment.

The source text does not report any illnesses, injuries, or deaths associated with this recall. Consumers and healthcare providers should stop using the affected devices and contact Boston Scientific for further instructions.

The recalled product

Product
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470
Affected units
382

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Material Number: M001195470
  • UDI-DI: 08714729839545
  • Batch Numbers: 38112474
  • 38149823
  • 38181962
  • 38181963
  • 38399282
  • 38415919
  • 38416360
  • 38445015
  • 38649840
  • 38649849
  • 39143221
  • 40470762
  • 40470763
  • 40470782
  • 40476387
  • 40476388
  • 40478984
  • 40478985

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Same hazard · manufacturing defect

See all →