ICU Medical Recalls 128-inch Primary Sets Due to Seal Defect
ICU Medical is recalling specific lots of 128-inch primary sets because a manufacturing defect may prevent the NanoClave spike from sealing properly.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk-of-harm product (medical device) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
ICU Medical, Inc. is recalling specific lots of 128-inch (325 cm) primary sets. The recall involves a potential manufacturing defect on the internal surface of the NanoClave component. This defect may inhibit a proper seal with the NanoClave spike.
The affected products are sterile, non-pyrogenic intravascular administration sets used to provide access for the administration of fluids from a container to a patient's vascular system. The recall covers a total of 304,735 units across lots 4711125 and 4718711.
The products were distributed nationwide in the United States and in several countries including Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia. Healthcare facilities should stop using the affected lots and contact ICU Medical for further instructions.
The recalled product
- Product
- 128" (325 cm) Appx 16.6 ml, 15 Drop Primary Set w/2 MicroClave, Remv 2 Gang 4-Way Clave Stopcock, Rotating Luer, 2 Ext. 1 unit per pouch, 50 multivac pouches per case.UDI:(01)10887709011691(17)250201(30)50(10)4711125;(01)10887709011691(17)250201(30)50(10)4718711. The Intravas
- Manufacturer
- ICU Medical, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Lot Numbers:4711125
- 4718711.
Distribution
Part of a larger recall action
This product is one of 99 recalled by ICU Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 99 products in this recall →Related recalls
Same manufacturer · ICU Medical, Inc.
See all →- HighICU Medical Recalls Hotline Fluid Warming Sets Due to Leakage Risk
FDA (Devices) · 2026-07-29
- SevereICU Medical Plum Duo Infusion Pump Software Defect Affecting Drug Delivery
FDA (Devices) · 2026-07-22
- SevereICU Medical Plum Solo Precision IV Pump May Deliver Excess Medication
FDA (Devices) · 2026-07-22
- HighPlum Duo Precision IV pumps may not display soft limit alerts
FDA (Devices) · 2026-06-17
- HighPlum Solo Precision IV pumps may not display soft limit alerts
FDA (Devices) · 2026-06-17
Same hazard · manufacturing defect
See all →- HighBeak & Skiff Alcohol Antiseptic Sanitizer recalled for acetaldehyde contamination
FDA (Drugs) · 2026-09-02
- HighSun Pharmaceutical Recalls Carbamazepine Tablets Due to Manufacturing Deviations
FDA (Drugs) · 2026-08-26
- HighKnee Prosthesis Trial Recall for Manufacturing Defect
FDA (Devices) · 2026-08-12
- HighU2 Total Knee System Femoral Valgus Adaptor Recall
FDA (Devices) · 2026-08-12
- HighUSTAR II Knee System Hinge Box Trial Recall
FDA (Devices) · 2026-08-12