The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0233-2027·Announced 2026-10-28

Boston Scientific Recalls Direxion Torqueable Microcatheters for Coating Defects

Boston Scientific is recalling 3,284 Direxion Torqueable Microcatheters due to insufficient UV curing of the hydrophilic coating and raised ridges on nitinol sleeves.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as Class II, and the source identifies manufacturing defects involving insufficient UV curing and laser cutting misalignment. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Boston Scientific Cork, Ltd. is recalling 3,284 units of the Direxion Torqueable Microcatheter, Material Number M001195300. The recall covers specific batch numbers and was issued by the U.S. Food and Drug Administration under recall number Z-0233-2027.

The product is defective due to insufficient UV curing of the hydrophilic coating on the outer surface of the distal segment of the microcatheters. Additionally, the devices have raised ridges on nitinol sleeves caused by laser cutting misalignment.

The affected devices were distributed worldwide, including nationwide in the United States and the territory of Puerto Rico, as well as in numerous other countries. The source text does not report any illnesses, injuries, or deaths associated with this recall.

The recalled product

Product
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300
Affected units
3,284

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Material Number: M001195300
  • UDI-DI: 08714729839415
  • Batch Numbers: 37818213
  • 37818214
  • 37818215
  • 37818216
  • 37881115
  • 37886786
  • 37886787
  • 37886788
  • 37914463
  • 37914464
  • 37914465
  • 37943140
  • 37956679
  • 37957740
  • 37977611
  • 38002444
  • 38002445
  • 38085044

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Same hazard · manufacturing defect

See all →

Similar recalls