The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0203-2027·Announced 2026-10-28

Medtronic CareLink Clinic software recall due to insulin calculation defect

Medtronic is recalling CareLink Clinic software version 4.4A because a defect in the Settings Calculator feature may result in incorrect rapid-acting insulin parameters.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The source describes a Class II recall for a software defect that can cause incorrect insulin dosing, with potential severity ranging from minor to catastrophic death. The source does not report any specific illnesses, injuries, or deaths.

Plain-English summary

Medtronic MiniMed, Inc. is recalling CareLink Clinic software version 4.4A. The recall is due to a software defect in the Settings Calculator feature where the calculation logic fails to use custom values entered by healthcare professionals for Total Daily Dose or Long-acting insulin. Instead, the calculator uses system-recommended values, resulting in incorrect rapid-acting insulin parameters for Carb Ratio, Insulin Sensitivity Factor, and Max Recommended value.

If these incorrect setting values are input into a patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their actual needs. This could lead to over or under delivery of insulin, with potential severity ranging from minor self-treatable hypoglycemia or hyperglycemia to catastrophic death from severe hypoglycemia.

The affected software was available in the United States between June 4, 2026, and June 8, 2026, with distribution in Kansas. Healthcare professionals and patients should verify their software version and consult Medtronic for guidance on correcting insulin parameters.

The recalled product

Product
CareLink Clinic, REF: MMT-7350
Affected units
1

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • 2026
  • and June 8th
  • 2026.

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically: