The Recall Desk
ModerateFDA (Devices)·Z-0126-2016·Announced 2015-10-21

Synthes Screw Inserter Recalled for Incorrect Etched Graphics

Synthes (USA) Products LLC is recalling 793 units of the Screw Inserter because the etched graphics may not correspond to the instrument's actual angle.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a situation where use of the product could cause temporary or medically reversible adverse health consequences. The defect is a labeling error (incorrect etched graphics) rather than a structural failure or contamination, and no injuries are reported in the source text.

Plain-English summary

Synthes (USA) Products LLC is recalling 793 units of the Synthes Screw Inserter, an orthopedic device used with Intramedullary Rods. The device is used to secure the head element to the head element inserter to prevent disengagement during insertion.

The recall is due to a labeling defect where certain lots of the Screw Inserter may be etched with incorrect graphics. Specifically, the orientation of the etched image of the oblique angle on the Screw Inserter may not correspond to the tip of the angle of the instrument.

The affected products are distributed nationwide in the United States. The recall involves Part Number 03.037.025 and specific lot numbers including 8911887, 8911975, 9166356, 9285766, 9286671, 9319408, 9375245, 9375251, 9388359, 9388360, 9393071, 9393074, 9393075, 9393076, 9409795, 9409796, 9422892, 9422895, 9498455, 9310899, 9310900, and 9319407. Healthcare providers should check their inventory for these specific part and lot numbers.

The recalled product

Product
Synthes Screw Inserter; for use with Intramedullary Rod, orthopedic device. Used to secure the head element to the head element inserter to prevent disengagement during head element insertion.
Affected units
793

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Part Number 03.037.025 Lot Numbers 8911887
  • 8911975
  • 9166356
  • 9285766
  • 9286671
  • 9319408
  • 9375245
  • 9375251
  • 9388359
  • 9388360
  • 9393071
  • 9393074
  • 9393075
  • 9393076
  • 9409795
  • 9409796
  • 9422892
  • 9422895
  • 9498455
  • 9310899

Distribution

Distributed nationwide across the United States.