Roche Recalls Tina-quant Complement C4 Reagents Due to Quality Issues
Roche Diagnostics is recalling 180 units of Tina-quant Complement C4 reagents due to cross-contamination causing invalid calibrations.
What this means for you
Cosmetic or documentation issue. No expected health impact, but the manufacturer is correcting the affected products.
Our severity reasoning: The recall is classified as FDA Class III, which the rubric defines as unlikely to cause adverse health consequences, typically resulting in a score of 1 or 2.
Plain-English summary
Roche Diagnostics Operations, Inc. is recalling 180 units of Tina-quant Complement C4 ver.2, Catalog 05991994190, lot 36870301. The products were distributed domestically nationwide.
The recall was initiated after Roche confirmed customer complaints regarding low quality control recovery and invalid calibrations. The company identified that a yellow discoloration was observed, which resulted from cross-contamination between the R1 and R3 reagents.
This recall is classified as Class III by the FDA, indicating that use of or exposure to the product is not likely to cause adverse health consequences. Users should stop using the affected lot and contact Roche Diagnostics for further instructions.
The recalled product
- Product
- Tina-quant Complement C4 ver.2, Catalog 05991994190
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Affected units
- 180
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- lot 36870301
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Roche Diagnostics Operations, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Roche Diagnostics Operations, Inc.
See all →- HighRoche cobas analyzer may repeat previous results for spline calibrated assays
FDA (Devices) · 2026-03-11
- HighThyroid receptor immunoassay recalled over variable results across instrument platforms
FDA (Devices) · 2026-01-14
- HighMedical Device Homocysteine Test Assay Recalled for Calibration Defects
FDA (Devices) · 2024-11-27
- HighRoche Creatine Kinase test reagents recalled for calibration errors
FDA (Devices) · 2024-10-02
- HighRoche Tina-quant D-Dimer Test System Recalled for Inaccurate Results
FDA (Devices) · 2023-11-29
Same hazard · cross contamination
See all →- CriticalCetirizine Hydrochloride Tablets Recalled Due to Cross Contamination with Ranitidine
FDA (Drugs) · 2026-07-29
- ModeratePrimidone tablets recalled over potential acemetacin cross-contamination at API manufacturer
FDA (Drugs) · 2026-06-17
- HighFly By Jing creamy sesame noodles may contain undeclared peanut
FDA (Food) · 2026-06-03
- LowPrimidone Tablets Recalled for Cross-Contamination with Acemetacin
FDA (Drugs) · 2026-05-27
- LowPrimidone Tablets 50 mg Recalled for API Cross-Contamination
FDA (Drugs) · 2026-05-27