Hospira Recalls GemStar Power Supply Due to Power Delivery Failure
Hospira is recalling GemStar Power Supplies because they may fail to provide direct current power to the infusion pump, potentially causing smoke.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device accessory with a reported instance of smoke, indicating a risk of harm (fire/burn) where injury has not yet been reported, fitting the 'High' severity criteria for risk-of-harm products.
Plain-English summary
Hospira Inc. is recalling the GemStar Power Supply, an accessory for the GemStar Infusion Pump. The power supply is a two-piece assembled switching unit that converts alternating current (AC) line input voltage of 120 or 240 V to an output of 3.3 Volts DC for the device.
The recall is being conducted because the GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one reported instance, smoke was observed while the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
The recall affects 4,709 power supplies distributed in the United States and 7,425 power supplies distributed in foreign countries. The products were distributed worldwide, including in the United States and various countries in Europe, South America, Africa, and the Middle East. Consumers and healthcare facilities should stop using the affected power supplies and contact Hospira Inc. for instructions on replacement or return.
The recalled product
- Product
- The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
- Manufacturer
- Hospira Inc.
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Desktop GemStar Power Supply List Number: 13072-05
- 1) US Distribution Lot Numbers: 10115G1
- 11121G1
- 11122G1
- 1123G1
- 11124G1
- 12143G1
- 12144G1
- 12178G1
- 15185G1
- 12195G1
- 17214G1
- 18217G1
- 18219G1
- 19225G1
- 20233G1
- 21242G1
- 22251G1
- 23259G1
- 24271G1
Distribution
Part of a larger recall action
This product is one of 2 recalled by Hospira Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Hospira Inc.
See all →- CriticalHospira Recalls Dextrose Injection Syringes Due to Human Hair Particulate
FDA (Drugs) · 2017-05-31
- ModerateHospira Recalls Minibore Extension Sets Due to Solvent Occlusion
FDA (Devices) · 2017-01-18
- LowHospira Recalls Lidocaine and Dextrose Injection for Stability Failures
FDA (Drugs) · 2016-12-14
- SevereHospira Recalls Fentanyl Citrate Injection Ampules Due to Broken Tips
FDA (Drugs) · 2016-11-30
- HighHospira Symbiq Infusion Pump Recall for White Screen Safety Shutdown
FDA (Devices) · 2016-11-16
Same hazard · fire
See all →- HighKibbi Recalls 2026-2027 Renegade RVs Due to Fire Risk
NHTSA · 2026-09-05
- SevereAmana Through-The-Wall Air Conditioners Recalled Due to Fire Risk
CPSC · 2026-09-03
- Severe
- High
- SevereTiffin Recalls 2026-2027 Wayfarer RVs Due to LP Tank Defects
NHTSA · 2026-08-31