The Recall Desk
ModerateFDA (Devices)·Z-0041-2027·Announced 2026-10-07

ASP Global Recalls Disposable Tourniquet Cuffs Due to Incorrect Expiration Dates

ASP Global is recalling sterile disposable tourniquet cuffs because they are labeled with a 10-year shelf life, but the correct validated shelf life is 5 years.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: This is a Class II recall involving a labeling error (incorrect shelf life) without reported injuries or illnesses, fitting the criteria for a moderate severity voluntary precautionary recall.

Plain-English summary

ASP Global, LLC. is recalling 15,580 units of the ASP-564008 Disposable Tourniquet Cuff, 24" x 4" 1-Line, Yellow. The recall is due to incorrect expiration dates on products manufactured between September 2016 and March 2023. The products are labeled with a 10-year shelf life, but the correct validated shelf life is 5 years.

The affected tourniquet cuffs are sterile, non-invasive devices intended to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. The distribution was nationwide across 34 states, including MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, and WY.

The specific lot codes affected are P1015STY, P1215STY, O0815STY, O1020STY, N0330STY, N0530STY, M0130STY, M0430STY, M0730STY, M1015STY, M1101STY, M1115STY, L0228STY, L0310STY, and L0325STY. The UDI-DI for the product is (01) 00840216106734. Healthcare facilities and consumers should check their inventory for these specific lot codes and stop using the affected products.

The recalled product

Product
Brand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 24" X 4" 1-Line, Yellow Model/Catalog Number: ASP-564008 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to…
Manufacturer
ASP Global, LLC.
Affected units
15,580

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (14)

  • P1215STY
  • O0815STY
  • O1020STY
  • N0330STY
  • N0530STY
  • M0130STY
  • M0430STY
  • M0730STY
  • M1015STY
  • M1101STY
  • M1115STY
  • L0228STY
  • L0310STY
  • L0325STY

Distribution

Distributed nationwide across the United States.

The notice also names 35 states specifically: