The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0039-2019·Announced 2018-10-10

Oridion Medical CAPNOSTREAM 20P Monitors Recalled for Settings Reset

Oridion Medical is recalling 264 CAPNOSTREAM 20P bedside patient monitors because date, time, and alarm settings may reset to factory defaults when powered off.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the defect involves the potential loss of critical alarm and nurse call settings on a medical device, which poses a risk of harm to patients even though no specific injuries are reported in the source text.

Plain-English summary

Oridion Medical 1987 Ltd. is recalling 264 units of the CAPNOSTREAM 20P (INTL) N - ROHS bedside patient monitors, Part Number CS08652-03. The recall is due to a defect where the date, time, nurse call, and alarm settings may reset to factory default settings when the monitor is powered off.

The affected products were distributed nationwide in the United States and Puerto Rico, as well as to numerous foreign countries including Argentina, Australia, Austria, Bahrain, Belgium, Canada, and others. The recall covers specific lot codes ranging from PA12170508 to PA43174149.

Healthcare facilities and consumers should check their inventory for the affected part number and lot codes. If a unit is affected, it should be removed from service and returned to Oridion Medical 1987 Ltd. for evaluation or replacement.

The recalled product

Product
CAPNOSTREAM 20P (INTL) N - ROHS, Part Number CS08652-03
Affected units
264

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Part of a larger recall action

This product is one of 37 recalled by Oridion Medical 1987 Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 37 products in this recall →

Similar recalls