The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0025-2027·Announced 2026-10-07

ASP Global Recalls Disposable Tourniquet Cuffs Due to Incorrect Expiration Date

ASP Global, LLC. is recalling 396 units of Expressaire Disposable Tourniquet Cuffs because the products are labeled with a 10-year shelf life, but the correct validated shelf life is 5 years.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, which typically indicates a temporary or medically reversible adverse health consequence. The hazard is a labeling error regarding shelf life, which fits the criteria for a moderate severity recall involving minor labeling errors or low-risk contamination.

Plain-English summary

ASP Global, LLC. is recalling 396 units of Expressaire Disposable Tourniquet Cuffs, Model A70107004. The recall is due to an incorrect expiration date on the product packaging. The products are labeled with a 10-year shelf life, but the correct validated shelf life is 5 years.

The affected units were manufactured between September 2016 and March 2023. The specific lot codes are R0310STY and R1201STY. The UDI-DI is (01) 00885914035481. These sterile disposable tourniquet cuffs are used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery.

The products were distributed nationwide in the United States, including in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, and WY. Healthcare facilities and consumers should stop using the affected products and contact ASP Global, LLC. for further instructions.

The recalled product

Product
Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 15" X 4" 1-LINE (CS 12) GRAY Model/Catalog Number: A70107004 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated…
Manufacturer
ASP Global, LLC.
Affected units
396

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • R1201STY

Distribution

Distributed nationwide across the United States.

The notice also names 35 states specifically:

Similar recalls