The Recall Desk

Manufacturer

Zeiss, Carl Inc

6 recalls in our database name Zeiss, Carl Inc as the manufacturing or recalling firm.

What the numbers show

Total
6
Critical
0
Severe
1
Most recent
2022-08-03

Of the 6 recalls from Zeiss, Carl Inc we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (17%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–6 of 6

  • LowFDA (Devices)·Z-1465-2022·2022-08-03

    ZEISS Axio Observer 5 Microscope Recalled for Incorrect Device Identification Label

    Zeiss is recalling ZEISS Axio Observer 5 microscopes due to an incorrect Unique Device Identification (UDI) label installed on some units. The labeling discrepancy affects device tracking and identification.

    Product
    ZEISS Axio Observer 5 microscope, Product material no. 431006-9901-000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1629-2020·2020-04-15

    Zeiss Laser Scanning Microscope Component Recalled for Laser Safety Bypass

    Zeiss is recalling a component for LSM 700, 800, and 900 microscopes because tilting the light arm can bypass laser safety, causing Class IIIB laser radiation emission.

    Product
    Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2441-2018·2018-07-25

    Zeiss UV Disinfection Unit Recalled for Potential Radiation Exposure

    Zeiss is recalling 8 UV disinfection units for the Celldiscoverer 7 microscope because a firmware issue may allow the unit to activate outside its housing, exposing users to harmful UV radiation.

    Product
    Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2058-2018·2018-06-20

    Zeiss Laser Scanning Microscopes Recalled for Potential Laser Shutter Failure

    Zeiss is recalling specific laser scanning microscopes because the laser shutter may not close completely, allowing reflected beams to exceed Class I safety limits.

    Product
    Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0799-2014·2014-01-29

    Zeiss Compensator Slider Lambda 6x20 Recalled for Misalignment

    Carl Zeiss Microscopy is recalling the Compensator Slider Lambda 6x20 component due to misalignment of the lambda plate.

    Product
    Compensator Slider Lambda 6x20 component (part number 473704-0000-000) utilized on all upright ZEISS microscopes (Axio Imager, Axio Scope. A1, Axio Lab. A1). Component used to determine composition of urinary calculi, urinary sediments, bio-crystallates and diagnosis of pathol
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1744-2013·2013-07-24

    Zeiss Microscope Polarizer D Component Recalled for Misaligned Plate

    Carl Zeiss Microscopy is recalling Polarizer D components for upright Zeiss microscopes due to a misaligned full-wave plate.

    Product
    Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means of color variation.
    Category
    Medical Device
    Distribution
    6 states