The Recall Desk

Manufacturer

Visionsense, Ltd.

3 recalls in our database name Visionsense, Ltd. as the manufacturing or recalling firm.

What the numbers show

Total
3
Critical
0
Severe
0
Most recent
2022-11-30

Of the 3 recalls from Visionsense, Ltd. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–3 of 3

  • HighFDA (Devices)·Z-0280-2023·2022-11-30

    Visionsense Infrared Microscope Recalled Due to Optical Fiber Cable Damage

    Visionsense is recalling 205 units of its VS3 Iridium microscope because optical fiber cables may become damaged during handling, causing loss of function and outer cable heating or melting.

    Product
    VS3 Iridium Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS-IR) Product Number: 174-0012
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0282-2023·2022-11-30

    VS3 Iridium Visionsense Beam Combiner Optical Fiber Cable Damage Risk

    Visionsense VS3 Iridium beam combiners may develop damaged optical fiber cables through handling or bending, causing loss of functionality and potential cable heating or melting. Affected units were distributed nationwide in the US.

    Product
    VS3 Iridium Visionsense Infrared (IR) Fluorescence Light Integrator (Beam Combiner) Product Number: 161-0001
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0281-2023·2022-11-30

    Medical device recall: VS3 Iridium infrared fluorescence microscope

    Visionsense, Ltd. issued a voluntary Class II recall of the VS3 Iridium infrared fluorescence microscope affecting 34 units with US Nationwide distribution.

    Product
    VS3 Iridium - Visionsense Infrared (IR) Fluorescence Miniature Microscope (MMS) - 785nm Product Number: 175-0012
    Category
    Medical Device
    Distribution
    Distributed nationwide