Siemens Healthcare Diagnostics, Inc.

111 recalls in our database name Siemens Healthcare Diagnostics, Inc. as the manufacturing or recalling firm.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

26–50 of 111

  • HighFDA (Devices)·Z-0597-2025·2024-12-11

    Siemens Atellica CI Analyzer may produce falsely elevated electrolyte test results

    Siemens' Atellica CI Analyzer may display incorrect diluent volume levels, causing falsely elevated sodium, potassium, and chloride test results in 559 units distributed worldwide.

    Product
    Atellica CI Analyzer. Catalog Numbers: 10947347.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0599-2025·2024-12-11

    Atellica CH β2-Microglobulin Reagent Produces Falsely Depressed Results

    Siemens Atellica CH β2-Microglobulin reagent may produce falsely depressed test results for samples between 18-30 mg/L, potentially leading to missed or delayed kidney disease diagnoses. No illnesses have been reported.

    Product
    Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0496-2025·2024-11-27

    Siemens Atellica CH Cuvette Segment manufacturing defect may impact lab test accuracy

    Manufacturing defects in Siemens Atellica CH Cuvette Segments may cause inaccurate laboratory test results for seven analytes across 9,729 units.

    Product
    Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0464-2025·2024-11-20

    IMMULITE 2000 PSA assay does not meet high-dose analytical performance claim

    Siemens Healthcare is recalling certain IMMULITE 2000 PSA assay diagnostic kits because they do not perform as claimed for high-concentration samples used on IMMULITE 2000 analyzers.

    Product
    IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3304-2024·2024-10-09

    ADVIA Centaur aTPO Assay kits recalled due to positive result bias

    Siemens Healthcare is recalling 5,622 ADVIA Centaur aTPO Assay kits due to test accuracy issues. The affected lots produce results that are biased higher at and below the 60 U/mL threshold.

    Product
    ADVIA Centaur aTPO Assay, Material Number 10630886 (100 Test) and 10630887 (500 Test)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3305-2024·2024-10-09

    Atellica IM aTPO Assay Diagnostic Test Recalled for Measurement Bias

    Siemens Healthcare Diagnostics is recalling 9783 kits of the Atellica IM aTPO Assay diagnostic test due to positive bias in patient results below the 60 U/mL cut-off level, which may produce inaccurate test results.

    Product
    Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3170-2024·2024-09-25

    Diagnostic test reagent may produce inaccurately low patient results

    Siemens Healthcare's ADVIA Chemistry Microalbumin diagnostic test may produce falsely low results. Eight lots do not meet High-Dose Hook Effect and Prozone Effect specifications, potentially affecting microalbumin detection accuracy.

    Product
    ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3169-2024·2024-09-25

    Siemens Atellica Microalbumin Diagnostic Test Kits Recalled for Measurement Error

    Siemens Healthcare Diagnostics is recalling Atellica CH and ADVIA Chemistry Microalbumin test kits worldwide because they may not properly measure microalbumin in urine samples, potentially causing falsely low results that could delay diagnosis of kidney disease.

    Product
    Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3065-2024·2024-09-18

    Siemens Dimension Vista Bilirubin Reagent Cartridges Recalled Due to Underfilled Wells

    Siemens Healthcare Diagnostics is recalling one lot of Dimension Vista Total Bilirubin Flex Reagent Cartridges due to underfilled wells during manufacturing. The defect may result in inaccurate test results for calibrators, quality control, and patient samples.

    Product
    Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3100-2024·2024-09-18

    Medical device diagnostic test shows measurement bias in certain samples

    Siemens Atellica IM CA 19-9 diagnostic tests show elevated measurement bias in certain sample ranges, particularly near the clinical threshold. Lot numbers 55974535 and 55975535 were distributed nationwide.

    Product
    Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995490
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3063-2024·2024-09-18

    Siemens Atellica CH diagnostic reagent batch may produce biased results

    A specific lot of Siemens Atellica CH urinary/cerebrospinal fluid protein reagent (lot 130414) may produce biased quality control results and inaccurate patient test outcomes. Approximately 229 units were distributed in the United States and 935 internationally.

    Product
    Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3104-2024·2024-09-18

    ADVIA Centaur CA 19-9 test kit recalled for measurement bias in samples

    Siemens Healthcare is recalling ADVIA Centaur CA 19-9 diagnostic test kits due to positive measurement bias in some samples. The affected kits may report falsely elevated test results, particularly around the normal limit of 35 U/mL.

    Product
    ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491244
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3102-2024·2024-09-18

    Siemens Atellica IM CA 19-9 Test Kit Recall: Measurement Bias in Samples

    Siemens is recalling Atellica IM CA 19-9 test kits (Lot 56434535) due to positive bias in measurement of tumor marker levels in some samples, particularly from Asia Pacific regions.

    Product
    Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 11206239
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3064-2024·2024-09-18

    ADVIA Chemistry urinary protein reagent quality control failure

    Siemens Healthcare is recalling ADVIA Chemistry urinary protein test reagents due to potential quality control issues that could produce inaccurate patient results.

    Product
    ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-3103-2024·2024-09-18

    Diagnostic test shows measurement bias at critical threshold

    Siemens Healthcare Diagnostics is recalling ADVIA Centaur CA 19-9 test kits due to positive bias in measurements around the critical threshold of 35 U/mL, affecting diagnostic accuracy.

    Product
    ADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491379 SMN: 11206239
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1886-2024·2024-05-29

    Thyroid Autoantibody Test System Recalled for Falsely Elevated Results

    Siemens IMMULITE 2000 Anti-TG Ab thyroid test kits are recalled due to the potential for falsely elevated patient sample results. Affected lots may produce inaccurate results that could lead to misdiagnosis.

    Product
    IMMULITE 2000 Anti-TG Ab Catalog # L2KTG2 (200 Tests), SMN 10381659, System, Test, Thyroid Autoantibody.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1887-2024·2024-05-29

    Siemens IMMULITE 2000 Anti-TG Ab Test Kits Recalled for Falsely Elevated Results

    Siemens Healthcare Diagnostics is recalling certain IMMULITE 2000 Anti-TG Ab test kits due to the potential for falsely elevated thyroid antibody results in patient samples.

    Product
    IMMULITE 2000 Anti-TG Ab Catalog # L2KTG6 (600 Tests), SMN 10381655, System, Test, Thyroid Autoantibody.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1549-2024·2024-04-24

    Siemens Atellica Lab Analyzers Risk Falsely Elevated Cholesterol Results

    Siemens Atellica CH and CI analyzers may produce falsely elevated cholesterol and lipid test results (2–16% high) after iron tests. This affects clinical calibration, quality control checks, and patient test results.

    Product
    Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1304-2024·2024-03-27

    Siemens DCA HbA1c Reagent Kit Recalled for Inaccurate Test Results

    Siemens is recalling its DCA HbA1c Reagent Kit because the device displays falsely elevated readings without alerting the user. This could lead to incorrect patient management of glucose levels.

    Product
    DCA Systems HbA1c Reagent Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1280-2024·2024-03-20

    Siemens Atellica IM Erythropoietin assay recalled for calibration bias error

    Siemens has recalled the Atellica IM Erythropoietin assay due to a 35% negative bias compared to WHO standards. The assay does not achieve claimed reference intervals and could report inaccurate test results.

    Product
    Atellica IM Erythropoietin (EPO) Assay (100 Test)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1281-2024·2024-03-20

    ADVIA Centaur Erythropoietin Assay Affected by -35% Measurement Bias

    Siemens Healthcare's ADVIA Centaur EPO Assay (lots 53984040, 55443042) produces readings 35% lower than the WHO standard, failing to achieve its claimed reference interval. No illnesses reported, but diagnostic inaccuracy poses risk of wrong clinical decisions.

    Product
    ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1003-2024·2024-02-14

    Siemens Atellica CH Immunoglobulin M Reagent Quality Control Bias

    Siemens is recalling 7,075 units of its Total Immunoglobulin M diagnostic reagent due to potential negative bias in quality control and patient sample results. This could affect the accuracy of test results.

    Product
    Total immunoglobulin M (IgM total) IVD, kit, nephelometry/turbidimetry
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2666-2023·2023-10-11

    Siemens ADVIA Chemistry Iron_2 Reagents Recalled for Potential Measurement Bias

    Siemens Healthcare Diagnostics is recalling ADVIA Chemistry Iron_2 Reagents due to a potential positive bias in test results that could affect the accuracy of Quality Control and patient iron level measurements.

    Product
    ADVIA Chemistry Iron_2 (IRON_2) Reagents, Material Numbers 10377510 (6 x 350 tests) and 10341118 (7 x 145 tests)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2472-2023·2023-09-06

    Siemens Atellica CH LDL Cholesterol analyzer reagent carryover issue

    Siemens is recalling Atellica CH analyzers due to potential reagent carryover that can cause falsely elevated results in cholesterol, bilirubin, and liver enzyme tests. This could affect the accuracy of patient laboratory results.

    Product
    Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code: LDLC Siemens Material Number (SMN): 11537214
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1271-2023·2023-09-06

    Atellica CH Triglycerides_2 Reagent Recalled Due to Magnesium Test Interference

    Siemens Healthcare Diagnostics is recalling 5,315 units of Atellica CH Triglycerides_2 diagnostic reagent due to reagent carryover that can affect magnesium test results. The product was distributed nationwide and internationally.

    Product
    Atellica CH Triglycerides_2: in vitro diagnostic use in the quantitative determination of triglycerides in human serum and plasma (lithium heparin, sodium heparin, potassium EDTA) using the Atellica CH Analyzer. SMN: 11537222
    Category
    Medical Device
    Distribution
    Distributed nationwide