The Recall Desk

Manufacturer

RAYSEARCH LABORATORIES AB

54 recalls in our database name RAYSEARCH LABORATORIES AB as the manufacturing or recalling firm.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

26–50 of 54

  • HighFDA (Devices)·Z-1194-2025·2025-02-26

    Medical device recall: RayStation radiation therapy software dose calculation inconsistency

    RayStation radiation therapy treatment planning software contains an inconsistency in density uncertainty calculations that may affect dose accuracy for proton and light ion treatments. The issue affects systems using HU-to-mass density CT calibration curves from RayStation versions 4.5 through 2024B.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 5.0.2.35 and 5.0.3.17 Software Version: RayStation 5 Service Pack 2 and 3
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1191-2025·2025-02-26

    RayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency

    RayStation radiation therapy planning software contains an inconsistency in how it calculates density uncertainty for proton and light ion treatments. This affects versions 4.5 through 2024B and could impact treatment plan accuracy.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1190-2025·2025-02-26

    RayStation radiation therapy software: incorrect density uncertainty calculations in dose planning

    RayStation radiation therapy planning software versions 4.5 through 2024B contains an inconsistency in density uncertainty calculations affecting proton and light ion dose computations. No illnesses reported.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1185-2025·2025-02-26

    Radiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations

    RayStation radiation therapy planning software contains an inconsistency in density uncertainty calculations within dose planning functions affecting proton and light ion treatments in versions 4.5 through 2024B.

    Product
    Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1188-2025·2025-02-26

    Medical Device RayStation Dosage Calculation Inconsistency in Radiation Therapy Planning

    RayStation radiation therapy treatment planning software contains an inconsistency in how it handles density uncertainty during dose calculations for proton and light ion treatments, potentially affecting treatment planning accuracy.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1193-2025·2025-02-26

    RayStation radiation therapy software recall for density calculation inconsistency

    RayStation treatment planning software versions 4.5-2024B contain a calculation error in density uncertainty functions affecting proton and light ion treatment plans. The error could lead to incorrect dose calculations.

    Product
    Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1741-2024·2024-05-08

    RayStation 7.0.0.19 radiation therapy planning software recalled for dose calculation error

    RayStation 7.0.0.19, a radiation therapy treatment planning system, is being recalled due to potential source-to-surface distance (SSD) calculation errors that could affect radiation dose accuracy. The recall affects 14 units distributed worldwide.

    Product
    RayStation 7.0.0.19. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2024·2024-05-08

    Radiation therapy planning system doses may be calculated incorrectly

    Certain versions of RayStation radiation therapy planning software may incorrectly report dose-related parameters, potentially affecting treatment calculations. The manufacturer recommends verification of affected systems.

    Product
    RayStation 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 &12.3.0.119. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1751-2024·2024-05-08

    RayStation radiation therapy planning system may report inaccurate dosimetry values

    RayStation 14.0.0.3338, a radiation therapy treatment planning system, may report source-to-surface distance (SSD) values as higher than actual measurements, potentially affecting radiation dose calculations. The FDA Class II recall involves 49 units distributed worldwide.

    Product
    RayStation 14.0.0.3338. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1752-2024·2024-05-08

    RayStation radiation therapy system dosing parameter recall

    RayStation 15.0.0.430 radiation therapy planning system may have an elevated SSD parameter. Users should contact Raysearch Laboratories AB for corrective guidance.

    Product
    RayStation 15.0.0.430. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1743-2024·2024-05-08

    RayStation radiation therapy planning software affected by source-to-surface distance calculation error

    RayStation radiation therapy treatment planning software versions 8.1.0.47, 8.1.1.8, and 8.1.2.5 may calculate source-to-surface distance incorrectly, potentially affecting dose calculations. Approximately 5 units are distributed worldwide.

    Product
    RayStation 8.1.0.47, 8.1.1.8 and 8.1.2.5. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1747-2024·2024-05-08

    Radiation therapy planning system may calculate patient doses too high

    RayStation versions 10.1.0.613 and 10.1.1.54 may calculate radiation doses higher than intended. Eighteen units have been distributed worldwide to medical facilities.

    Product
    RayStation 10.1.0.613 and 10.1.1.54. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1744-2024·2024-05-08

    Radiation Therapy Planning System May Report Dose Values Too High

    RayStation radiation therapy planning software versions 9.0.0.113 and 9.0.1.142 may report dose values that are higher than actual. The Class II recall affects 63 units distributed worldwide.

    Product
    RayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1745-2024·2024-05-08

    Radiation therapy planning system dose calculation error

    RayStation versions 9.1.0.933 and 9.2.0.483 may incorrectly calculate dose parameters. 11 units of this radiation therapy treatment planning software were distributed worldwide.

    Product
    RayStation 9.1.0.933 and 9.2.0.483. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1750-2024·2024-05-08

    RayStation Radiation Therapy System Recalled for Potential Dosing Error

    RayStation versions 13.0.0.1547, 13.1.0.144, and 13.1.1.89 radiation therapy treatment planning systems are recalled due to a potential software error in source-to-surface distance calculations that could affect radiation dose delivery.

    Product
    RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1746-2024·2024-05-08

    RayStation Radiation Therapy Planning Software Recalled for Calculation Error

    RayStation radiation therapy planning software versions 10.0.0.1154, 10.0.1.52, and 10.0.2.10 have been recalled due to a potential calculation parameter issue that could affect treatment planning. The FDA classified this as a Class II recall affecting 213 units worldwide.

    Product
    RayStation 10.0.0.1154, 10.0.1.52 & 10.0.2.10. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1742-2024·2024-05-08

    Radiation therapy planning system may report overly high skin surface doses

    RayStation versions 8.0.0.61 and 8.0.1.10 are subject to a Class II FDA recall due to a potential for calculated skin surface doses to be reported as too high. Healthcare facilities using these versions should contact the manufacturer for guidance and a corrective update.

    Product
    RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2024·2024-05-08

    Radiation Therapy Planning System May Calculate Skin Surface Distance Incorrectly

    RayStation radiation therapy planning software versions 11.0.0.951, 11.0.1.29, 11.0.3.116, and 11.0.4.15 may calculate skin surface distance (SSD) values incorrectly, potentially affecting dose calculations. Affected software has been distributed worldwide.

    Product
    RayStation 11.0.0.951, 11.0.1.29, 11.0.3.116 & 11.0.4.15. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1482-2024·2024-04-17

    RayStation Software Use Error Could Cause Incorrect Radiation Dose Calculations

    RayStation radiation therapy planning software versions 8B through 2024A contain a use error that can incorrectly select silicon material for silicone gel polymer implant dose calculations, potentially resulting in incorrect radiation doses.

    Product
    RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0206-2024·2023-11-08

    Oncology Information System Document Values Not Visible in Read-Only Mode

    RayCare oncology information system versions 5B and 6A contain a software bug where document field values do not display when documents are opened in read-only mode outside the Documents workspace.

    Product
    RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-2464-2023·2023-09-06

    Oncology Software Allergy Warnings May Not Display for Medication Ingredients

    RayCare oncology software versions 5A, 5B, and 6A may fail to display medication allergy warnings under certain circumstances. Healthcare facilities could inadvertently prescribe medications patients are allergic to.

    Product
    RayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. Model Numbers: 5.0.0, 5.0.1, 5.1.1,5.1.2,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1281-2023·2023-03-29

    Raystation Treatment Planning System Collimator Angle Export Defect

    Raystation radiation therapy planning software may not correctly export collimator angle settings when using the DICOM export function. In exported files, the collimator angle may be replaced with zero regardless of the user's selection.

    Product
    Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B,
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1403-2022·2022-07-27

    Radiation treatment planning software fails to propagate treatment course data

    RayStation 9B SP1 radiation treatment planning software has an issue where treatment course information may not be propagated to other systems in some workflows. This could potentially affect treatment planning accuracy.

    Product
    RayStation 9B SP1. For radiation treatment planning.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1299-2022·2022-07-06

    Radiation therapy software shows misleading dose display in RayStation 11B

    RayStation 11B radiation therapy planning software has two display issues affecting Linear Energy Transfer values. These errors could lead to incorrect dose calculations if not identified during clinical review.

    Product
    RayStation 11B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0849-2022·2022-04-06

    Radiation therapy planning software: incorrect dose density calculation

    RayStation/RayPlan radiation therapy software (versions 4–11A) may incorrectly calculate radiation dose density in treatment planning. The error affects dose calculations in regions partially covered by certain structures.

    Product
    RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47,
    Category
    Medical Device
    Distribution
    Distributed nationwide