The Recall Desk

Manufacturer

Primus Medical LLC

3 recalls in our database name Primus Medical LLC as the manufacturing or recalling firm.

What the numbers show

Total
3
Critical
0
Severe
1
Most recent
2014-10-22

Of the 3 recalls from Primus Medical LLC we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (33%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–3 of 3

  • HighFDA (Devices)·Z-0067-2015·2014-10-22

    Primus Medical Recalls AC Powered Adjustable Bed Frames

    Primus Medical LLC is recalling 1,726 AC powered adjustable bed frames used in nursing facilities because the head deck actuator bracket may separate from the frame.

    Product
    AC Powered adjustable bed frame. AC powered adjustable bed frame used in nursing facilities.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2596-2014·2014-09-17

    Primus Medical Recalls Transcend and Bariatric Mattresses Due to Delamination

    Primus Medical LLC is recalling specific Transcend and Bariatric mattresses because the top cover is delaminating. The recall affects 1,557 units distributed in the US and Canada.

    Product
    Prime Care ¿ Transcend Mattress, Prime Care ¿ Transcend Perimeter Mattress, Prime Care ¿ Transcend Advanced Mattress, Prime Care ¿ Transcend Advanced Perimeter Mattress, Prime Plus ¿ Bariatric Mattress and Prime Plus ¿ Bariatric Perimeter Mattress, individual units packaged seper
    Category
    Medical Device
    Distribution
    18 states
  • SevereFDA (Devices)·Z-0270-2013·2012-11-14

    Primus Medical Recalls Prime Bed TL500 Hospital Beds Due to Frame Collapse Risk

    Primus Medical LLC is recalling 342 Prime Bed TL500 hospital beds because retaining washers and clevis pins may fail, causing the bed frame to collapse.

    Product
    AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
    Category
    Medical Device
    Distribution
    Distributed nationwide